ecocool 100R
Page 11
Operating Manual
www.grantinstruments.com
Mains power outlet for accessory pump
Optional power connection for an accessory pump. Power can be set ON using the
Settings menu. See section 5.2.8
Mains power cable inlet
For connecting the equipment to a mains power supply. Use the mains cord supplied
with the system. See Section 4.6 for instructions
Top plate thumbscrew
To remove the heating circulator from the refrigerated bath for cleaning, first unfasten this
thumbscrew. Slide the top plate backwards and lift to completely remove the heating
circulator.
4.3
Installation
After transportation, let the equipment stand in its intended working position for six hours.
This is to allow the oil to drain to the bottom of the compressor. This is normal procedure
for refrigeration compressors.
Allow at least 100mm clearance from obstructions at the front and rear so that there is
free air flow through the equipment.
Place the equipment on a firm, level surface. Ensure that the mains plug and
the switch at the rear of the equipment are easily accessible.
After transport or storage in humid conditions, always allow the equipment to
stand for at least an hour at room temperature before operating.
Do not block or restrict ventilation slots. Allow at least 100mm clearance from
obstructions
Do not connect to a power supply or switch on before filling the tank.
Drain before moving the bath. Liquid to be drained should not be below 10°C
or above 50°C.
Always operate the equipment with the condenser grille fitted.
4.4
Rotating the heating circulator
The heating circulator is fitted ready to use with the pump ports facing the left-hand side.
If required, the heating circulator can be rotated so the pump ports face the rear and the
user controls are accessed from the left hand side.
Consider rotating the heating circulator, if following a risk assessment, access to hot
surfaces needs to be reduced.
To rotate, remove the thumbscrew at the rear of the refrigerated bath and lift up the
heating circulator and top plate. Remove the 2 thumbscrews from the underside of the
top plate to release the heating circulator. Rotate the heating circulator on the top plate
and refit.