Recirculating Chiller 

Part No 09931/Issue 

10

                                                 Page 

6

April

Rear view: RC3000G 

Wheels

Serial number label

3RZHUVXSSO\YROWDJHFRQQHFWRU

Remote alarm socket

Brakes

Brakes on front wheels.
Push down to lock.

Overflow pipe

Inlet  pipe 3/8" BSP
male
th

d

Outlet  pipe 3/8" BSP
male

                                                                         Figure 2 

 2.1  

Connection for closed loop circulation 

Use pipe of a suitable bore, which will withstand a pressure up to 40psi.  

 

The IN-let and OUT-let connections (3/8” BSP Males) are on the rear panel of the chiller (see 
figures 1 and 2). Connect the unit to the external apparatus; ensure that the connections are 
liquid tight. The RC should be as close as possible to the apparatus.

Connection for closed loop circulation with PRES Reservoir 

 

The PRES Reservoir can simplify the priming of systems when operating a Closed Loop system 
that has no filling port available on the suction side of the pump 

Attach the PRES reservoir to the rear panel of the circulator according to the instructions 
provided.  Remove the 3/8” BSP straight connector from the IN-let pipe. Screw one end of the 
flexihose assembly onto the IN-let pipe, then connect to the reservoir by holding the pipe in 
position whilst turning the other swivel connector on the reservoir on to the other stationary end of 
the flexihose. IF YOU ATTEMPT TO CONNECT IT ANY OTHER WAY THIS WILL KINK THE 
HOSE.

2.2       Connection for open loop circulation 

When circulating through an open tank, connect the RC inlet to the lowest point of the tank. The 
RC outlet hose should be positioned as far as possible from the RC inlet hose. Unless the set 
temperature of the liquid in the tank is the equal to the ambient temperature, insulate the tank. 

Summary of Contents for CIR2020

Page 1: ...Recirculating Chiller RC350G RC400G RC1400G RC3000G Operating Manual...

Page 2: ...Recirculating Chiller Part No 09931 Issue 10 Page 2 April 2010...

Page 3: ...ng 9 3 3 Over undertemperature alarm setting 9 3 3 1 Setting high alarm offset 10 3 3 2 Setting low alarm offset 10 3 4 Cooling 10 3 5 Tuning 10 3 6 Undertemperature thermostat 10 4 Safety devices 11...

Page 4: ...on slots x If liquid is spilt inside the unit disconnect it from the power supply and have it checked by a competent person x It is the users responsibilty to carry out appropriate decontamination if...

Page 5: ...te probe PRES Reservoir accessary pack PRES Reservoir 2 x Clips 2 x 3 8 to 1 2 adaptor 4 x M4 CSK screws Mounting instruction drawing Optional Remote Probe Rear view RC350G 400G 1400G chiller Wheels S...

Page 6: ...can simplify the priming of systems when operating a Closed Loop system that has no filling port available on the suction side of the pump Attach the PRES reservoir to the rear panel of the circulator...

Page 7: ...rature equal to the actual temperature Refer to section 3 2 for temperature setting In a closed loop system to prevent excessive pressures occurring when the liquid circulates system allowances for th...

Page 8: ...attachment of an external alarm device a light or buzzer for example The contact ratings are 2A at 240V Connect the external alarm socket pins as follows Pin1 Ground Pin 2 Output Pin 3 Output 2 7 Remo...

Page 9: ...this is highly energy efficient but means that the RC3000G can not be used for heating the most stable performance is achieved when cooling a large volume with a stable heat load 3 3 Over and undertem...

Page 10: ...g the initial commissioning of the process However if the process under control subsequently becomes unstable because its characteristics have changed you can retune again at any time for the new cond...

Page 11: ...en the flow starts 4 3 Overtemperature cut out This prevents the heater chamber from overheating If the temperature rises above 63 C the heater switches off and the alarms operate 4 4 High pressure cu...

Page 12: ...start contact Grant s service department RC cannot heat to set temperature Cooler switched on Switch cooler off If this has no effect see below External cooling load too large for RC There is no recov...

Page 13: ...Supply voltage 220 240V 50Hz RC350G RC400G RC1400G Class A RC3000G Temperature range 5 C to 60 C 10 C to 60 C 10 C to 60 C 10 C to 60 C Stability DIN 58966 20 C with 10L of water at 20 C with 25L of w...

Page 14: ...Replacement of fuses Fuse replacement should only be carried out by a competent person 7 2 1 RC350G and RC400G Disconnect the unit from the power supply Remove the IEC plug from the socket in the bac...

Page 15: ......

Page 16: ...Grant Instruments Cambridge Ltd Shepreth Cambridgeshire SG8 6GB England Tel 44 0 1763 260811 Fax 44 0 1763 262410 Email labsales grant co uk www grant co uk...

Reviews: