CALIBRATION

1.  Press 

, input password “13386” and press 

.

2.  It displays “d   *”, * means division value, you can choose 1, 2, 5, 10, 20, 

50 and then press 

.

3.  It displays “P   *”, * means decimal points of division value, choose 0 for 

no decimal or 1 means 1 decimal; then press 

.

4.  It displays Max capacity like “05000”, input Max Cap you want and press 

.

5.  It displays “noload”, wait until stable light on and press 

.

6.  It displays f.S., load weights onto the scale (like 3000kg), input value of 

weights like “03000”, wait until the stable light on, then press 

.

7.  Finished.

Summary of Contents for C6 SERIES

Page 1: ...SERIES SERIE C6 Indicador Impresora Indicator Printer OPERATION MANUAL MANUAL DE UTILIZACION...

Page 2: ...6 Statistics Printing 7 General Printing 7 Auto Printing 8 Division Value 8 Backlight 8 Accumulation 8 Accumulation check 9 Pre set Tare 9 Parameters setting 9 Zero and baud rate setup 9 Printing para...

Page 3: ...m tica 18 Valor de la division 18 Iluminaci n del display 18 Acumulaci n 19 Verificaci n de las acumulaciones 19 Predeterminaci n de la Tara 19 Configuraci n de par metros 19 Configuraci n del cero u...

Page 4: ...the battery power level before after use and recharge the scale in time In order to maximize the battery life periodic recharging is necessary for scales even not in use for long time Do not try to r...

Page 5: ...minute Sensitive 2 02 0 004mV V Supply voltage 3 7V DC 4 Voltage 7 4V DC 5 Power Approximate 0 2W 6 Operation temperature weighing display unit between 0 C to 40 C sensor 20 C 40 C r 7 Relative humid...

Page 6: ...122530 then press The time has been adjusted Date In weighing status if check date press key it displays current date like 11 03 10 Date adjustment Press key input 17 then press it displays date like...

Page 7: ...weighing the goods after finishing weighing all the goods press key accumulate all the weights and print out then quit the mode Sample General Printing When weighing goods after the stable light on p...

Page 8: ...ss it displays P 0 means no decimal point of division value e g to set division 0 5kg input 1 P 1 then press it displays d 1 input 5 d 5 then press Finished now the division value is 0 5kg Backlight P...

Page 9: ...parameter 3 The display will show F2 2 Zero function range of power on choose 0 means 0 1 means 4 2 means 50 3 means 100 Select the desired option and press to confirm and go to next parameter 4 The d...

Page 10: ...me of auto printing after stable light on choose 0 means the time after stable 1means 1s after stable 2 means 2s and the longs time is 3s press 5 The display will show END press to back to normal weig...

Page 11: ...t the value and press it ll begin printing 5 Delete data Press and input 72851 press then all the data will be deleted Setting up channel 1 Press key it displays P 00 Input 20 and press 2 The display...

Page 12: ...cimal points of division value choose 0 for no decimal or 1 means 1 decimal then press 4 It displays Max capacity like 05000 input Max Cap you want and press 5 It displays noload wait until stable lig...

Page 13: ...fect of the material During this period GRAM PRECISION SL covers the manpower and the spare parts necessary for the reparation of the scale This guarantee does not cover the failures caused by an inap...

Page 14: ...tes y despu s de la recarga Con el fin de maximizar la duraci n de la bater a se recomienda la recarga peri dica aunque el gancho se utilice durante un largo periodo No intentar reparar la balanza con...

Page 15: ...V V Tensi n de alimentaci n 3 7V DC 4 Voltaje 7 4V DC 5 Alimentaci n 0 2W aprox 6 Temperatura de trabajo 0 C a 40 C c lula de carga 20 C 40 C r 7 Humedad relative 10 90 R 8 Distancia entre la unidad d...

Page 16: ...la tecla introducir el c digo 17 y pulsar la tecla el display mostrar la fecha 110310 introducir la fecha correcta y pulsar la tecla para confirmar y guardar N mero de operaci n Si desea verificar el...

Page 17: ...ci n de todas las pesadas realizadas muestra Impresi n general Para imprimir los resultados de una pesada esperar a que lectura del display est estable y pulsar la tecla el indicador imprimir el n mer...

Page 18: ...division pulsar la tecla el display mostrar P 0 que significa sin punto decimal en el valor de la divisi n Por ejemplo si se desea configurar la divisi n a 0 5kg introducir P 1 y pulsar la tecla el di...

Page 19: ...i n del cero u la velocidad de transmisi n 1 Pulsarlatecla paraaccederalmenudeconfiguraci ndepar metros e introducir el c digo 18 pulsar la tecla 2 El display mostrar F1 1 Configuraci n del rango de c...

Page 20: ...deseado y pulsar para confirmar 7 El display mostrar END pulsar la tecla para volver al modo normal de pesaje Configuraci n de los par metros de impresi n 1 Pulsar la tecla e introducir el c digo 97...

Page 21: ...significa modo simple solo imprimir el peso total de los articulos 1 significa la impresi n de los detalles de los articulos Introducir el valor y pulsar para empezar a imprimir 3 Introducir 2 el dis...

Page 22: ...onfiguraci n de la auto desconexi n 1 Pulsar la tecla el display mostrar P 00 introducir el c digo 30 y pulsar la tecla 2 Eldisplaymostrar dl 01 01significaautodesconexi nen30minutos 02 significa auto...

Page 23: ...m xima 05000 introducir la capacidad m xima deseada y pulsar la tecla 5 El display mostrar noload esperar hasta el s mbolo de estabilidad se active y pulsar la tecla 6 El display mostrar F S colgar l...

Page 24: ...riodo de 1 a o a partir de la fecha de entrega Durante este periodo GRAM PRECISION SL se har cargo de la reparaci n de la balanza Esta garant a no incluye los da os ocasionados por uso indebido o sobr...

Page 25: ...25 Notes...

Page 26: ...26 Notes...

Page 27: ...27 Notes...

Page 28: ...Gram Precision S L Traves a Industrial 11 08907 Hospitalet de Llobregat Barcelona Spain Tel 34 902 206 000 34 93 300 33 32 Fax 34 93 300 66 98 comercial gram es www gram es 001 08072016...

Reviews: