Battery use and care:

 Use two AA alkaline batteries only. If the otoscope will not be used for an extended 

period of time, remove the batteries. Dispose of used batteries in accordance with all applicable laws and 
regulations.

Cleaning and sterilization:

 Dampen a clean cloth with a mild detergent and water solution, a solution 

of 70% isopropyl alcohol and water solution, or a 10% bleach and water solution. Wring out the cloth to 
remove excess fluid. Use the cloth to wipe all external surfaces of the otoscope. Ensure that no fluid enters 
the otoscope. Clean the instrument glass with a cotton swab dipped in isopropyl alcohol. The reusable ear 
specula may be sterilized in 134°C saturation steam for a dwell time of ten minutes.

s

  notice: do not immerse this device in fluid.

speciFicAtions

Optical specifications

Illumination form

FiberOptics transmit light

Lens magnification

2.5x

Otoscope port

View tympanic membrane using the Fiberoptic Otoscope bulb

Ear specula specifications

Ear specula diameter

2.5mm, 4mm

Light port specifications

Light port

Yes. An otoscope bulb can be fitted to the port.

Electrical specifications

Bulb model

T4954 2.5V, 1.6W

Rated voltage

Two AA alkaline batteries (not included)

Operation mode

ON time

Intermittent; ≤ one minute

OFF time

≥ five minutes

Electrical safety standard

IEC60601-1, IEC60601-1-2

Environmental requirements

Use

Temperature

50°F — 95°F (10°C — 35°C)

Relative humidity

30% — 75% RH

Atmospheric pressure

10.15 psi — 15.37 psi (70 kPa — 106 kPa)

Shipping and storage

Temperature

-40°F — 131°F (-40°C — 55°C)

Relative humidity

10% — 90% RH

Atmospheric pressure

7.25 psi — 15.37 psi (50 kPa — 106 kPa)

Replacement parts

 Specula size 2.5mm (8617-1);         Specula size 4.0mm (8617-2);         Replacement bulb (8617-3)

limited WArrAnty

GF Health Products, Inc. ("Graham-Field") warrants the Grafco 8617 Fiberoptic Otoscope (not including 
specula) for defects in workmanship and materials for a period of one year. If a product is deemed to be 
under warranty, GF Health Products, Inc. shall provide, at its option, (1) replacement of any defective part 
or product or (2) a credit of the original selling price made to GF Health Products, Inc.’s initial customer. 
The warranty does not include any labor charges incurred in replacement part(s) installation or any 
associated freight or shipping charges to GF Health Products, Inc.
The warranties contained herein contain all the representations and warranties with respect to the subject 
matter of this document, and supersede all prior negotiations, agreements and understandings with respect 
thereto. The recipient of this document hereby acknowledges and represents that it has not relied on any 
representation, assertion, guarantee, warranty, collateral contract or other assurance, except those set out 
in this document.

Reviews: