Meter Setup and Operation Instructions
3A1451C
9
Setup Menus (F
1.
If the display is blank (asleep), wake it up by press-
ing any button on the keypad (F
2.
Press and hold the Total and
Reset buttons simultaneously
for approximately six (6) sec-
onds to enter the Setup
Menus (F
. 10).
NOTE: During the first 4 seconds
all segments display. Then for 2 seconds the software
version number displays.
3.
There are three (3) Setup Menus available, stored in
a preset order.
a.
The first screen displayed is the Resettable
Total Menu. When this menu is displayed,
R-TOTAL will blink on the screen (F
IG
. 11).
b.
Press and hold Reset but-
ton to display Units of
Measure Menu. When
this menu is displayed the
last set Unit of Measure-
ment blinks on the screen
c.
Press and hold Reset button again to display
Calibration Menu. When this menu is displayed
CAL blinks on the screen (F
The total that is displayed when you leave each
menu is the total that is stored.
NOTE: The following sections of this manual provide
instructions for using the Setup Menus.
Resettable Total (F
IG
. 11)
Resets dispensed total on screen to zero or stores dis-
played dispense total. The resettable total accumulates
until the next time it is manually reset.
1.
If Resettable Total Menu is not already displayed,
complete steps 1 and 2 of Setup Menus section.
2.
R-TOTAL blinks indicating you are on the Resetta-
ble Total Menu. The currently stored total and unit of
measurement is displayed (F
IG
. 11).
3.
Do ONE of the following.
a.
Press and hold the Total
button. Resettable total is
set to zero (0).
0
GAL/QTS/PTS or 0
LITERS is displayed.
b.
Press and hold the Reset
button. The zero (0) total
is stored. Units of Mea-
surement Menu displays.
OR
F
IG
. 10
0D[LPXP:RUNLQJ3UHVVXUHEDU36,
*$/
476
376
/,7(56
5727$/
$872
ti12044a
F
IG
. 11
*$/
476
376
/,7(56
5727$/
$XWR
0D[LPXP:RUNLQJ3UHVVXUHEDU36,
ti12047a