GPI TM050 User Manual Download Page 5

5

TM Series Meters

  ½ inch meter 

1 GPM (3.8 LPM)

  ¾ inch meter 

2 GPM (7.5 LPM)

  1 inch meter 

5 GPM (18.8 LPM)

  1 ½ inch meter  

10 GPM (37.5 LPM)

  2 inch meter  

20 GPM (75 LPM)

The use of a uniformly dependable, accurate 

calibration container is recommended for the 

most accurate results. Due to high flowrate, it 

is strongly recommended that calibration be 

completed with a combination of volume and 

weight using fine resolution scales.
For best results, the meter should be installed 

and purged of air before field calibration.

Field Calibration with  

Computer Display

Field Calibration and Factory Calibration are 

defined in the Operation Section. Factory 

calibration settings are programmed into each 

computer during manufacturing, using water 

at 70° F (21° C). Readings using the Factory 

Calibration (FAC) may not be accurate in some 

situations, for example, under extreme tem-

perature conditions, non-standard plumbing 

configurations or with fluids other than water.

Field Calibration Procedures 

(Correction Factor Method)

1.  To calibrate, press and hold the CALIBRATE 

and DISPLAY buttons for about 3 seconds 

until you see FLdCAL. Release both buttons 

and you will see CF - 00.0. You are now in 

the field calibration mode and values from 

-99.9% to +99.9% can be entered.

2. The 

+

/– position appears either as an “under-

score” character for plus, or as a “hyphen” 

character for minus. The DISPLAY button 

selects the position and the CALIBRATE 

button toggles this character.

3.  The DISPLAY button can then be pushed 

to select the numeric positions. Press the 

CALIBRATE button to scroll from 0 to 9. 

Enter the percentage of change you want 

the display to correct. When satisfied with 

the value, press both CALIBRATE and DIS-

PLAY buttons simultaneously.  CALEnd will 

be displayed and unit will go back to normal 

operation, less the FAC (factory calibration) 

icon.

4.  All enabled units-of-measure remain visible 

and selectable – the entered correction will 

be applied to all enabled units.

5.  To return to factory calibration (FAC), press 

and hold both CALIBRATE and DISPLAY 

buttons for about 3 seconds until FAcCAL 

is displayed. Then release buttons. Unit 

should return to normal operation and FAC 

icon is visible.

Calibration with Conditioned  

Signal Output Module

The K-factor of your meter appears on the 

calibration report as the number of pulses per 

gallon. The factor is determined during produc-

tion using water at 70° F (21° C). This K-factor 

may be used for “single point” calibration and 

provide acceptable accuracy. However, readings 

may not be accurate when using this calibra-

tion method in some situations. For example, 

extreme temperature conditions, non-standard 

plumbing configurations or with fluids other 

than water.

MAINTENANCE

Proper handling and care will extend the life and 

service of the meter.

Turbine Rotor

The meter is virtually maintenance-free. How-

ever, it is important the rotor moves freely. Keep 

the meter clean and free of contaminants.
If the rotor does not turn freely, apply a pene-

trating lubricant on the rotor, shaft and bearings. 

Remove any debris or deposits from the rotor 

using a soft brush or small probe. Be careful 

not to damage the turbine rotor or supports.

Blowing compressed air through the tur-

bine assembly could damage the rotor.

 

CAUTION

 

CAUTION

 

CAUTION

Do not allow liquids to dry inside the turbine.

Handle the rotor carefully.  Small scratches 

or nicks can affect accuracy.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for TM050

Page 1: ...inches of some electric motors or other sources of electronic noise INSTALLATION Connections Install your meter in line either horizontally or vertically or at the end of the hose adjacent to the nozz...

Page 2: ...curls can adversely affect accu racy Correct the problem prior to turbine installation NOTE Do not over tighten the flange bolts This may cause the gasket to be compressed into the flow stream and ma...

Page 3: ...ule and lift the module from the turbine 2 To change terminal block connections loosen the appropriate screws Reconnect the wires in the proper positions andtighten the screws 3 Install the batteries...

Page 4: ...eplaceable batteries When display becomes dim faded or the low battery message appears see below the batteries need to be replaced Reference the Maintenance Section for details Factory and Field Calib...

Page 5: ...ct When satisfied with the value press both CALIBRATE and DIS PLAY buttons simultaneously CALEnd will be displayed and unit will go back to normal operation less the FAC factory calibration icon 4 All...

Page 6: ...becomes dim blank or the low battery message appears see below replace the batteries as follows 1 Remove the four Phillips head screws from the face of the meter and lift the faceplate from the turbin...

Page 7: ...ch 222 kg 381 kg 381 kg 1 inch 299 kg 626 kg 666 kg 2 inch 354 kg 807 kg 875 kg Dimensions cm W x H x L Without Fitting With NPT Fitting 5 3 x 6 4 x 10 9 5 3 x 6 9 x 15 2 5 3 x 6 9 x 11 2 5 3 x 7 3 x...

Reviews: