Device Description HG G-19603ZA | English, Revision 03 | Date: 31.03.2021
30
Chapter 6 – Software
If the compensation for a static magnetic field is activated the raw data shown in this
menu is already calculated with the compensation included (see section 4.2 on page
17).
6.4.6 Firmware Update
Start the service program (see section 6.4 on page 25).
Use the setting menu (s. section 6.4.3 on page 28) to put the device into the
DFU Modus (Device Firmware Upgrade) for the firmware update.
Figure 23
Menu firmware update
Close the connection via the COM port in the terminal program (
disconnect
).
For the following steps you need the firmware as a
.dfu
file as well as the software
DfuSe
by ST Microelectronics. You can get the firmware file upon request from Göt
-
ting. The software
DfuSE
is available from the following address:
http://www.st.com/en/development-tools/stsw-stm32080.html
Download
DfuSE
, install the program and start it. It starts into the Demo GUI
Mode which is sufficient to carry out the firmware update.
Use
Choose
to select the
.dfu
firmware file provided by Götting KG.
Figure 24
Firmware Update – Choose file
If the status bar says
File correctly loaded
start the firmware update with
Upgrade
.
entering DFU bootloader...
start DFU application or repower device
Use
Choose
to select the .dfu file
As shown the
DFU mode
has
to be active