Minimum Permissible Limit Values for Insulation Resistance 
Relative to Insulation Test Voltage per EN 62446 

Note

If the limit value is fallen short of, 
< 1 M

 or < 0.5 M

 appears for R

ISO

.

Example

9.5

General Insulation Resistance Measurement

The 

PROFITEST PV

SUN

 test instrument is additionally equipped with 

system-independent insulation testing. This is helpful, for exam-
ple, when testing systems or supply lines when no modules have 
yet been installed. All insulations tests known to date can be per-
formed.

Warning!

The test object must be potential-free. Assure absence 
of voltage with the help of a suitable measuring instru-
ment, for example the METRAVOLT 12D+L voltage tester.

Note

General insulation measurement is executed between 
short circuited (–) and (

) terminals.

Connect the (–) and (

) terminals at the 

PROFITEST PV

SUN

 to the 

device under test.

Switch the 

PROFITEST PV

SUN

 test instrument on.

Selecting the Test Voltage

Press and hold the 

RISO/250V/500V/1000V 

key for at least 2 

seconds.

The currently selected test voltage appears at the display.

Select the required insulation test voltage by briefly and re-
peatedly pressing the 

R

ISO

/250V/500V/1000V 

key.

The last selected value is saved after a short period of time, and is 
then cleared from the display.

Example

Executing the Insulation Resistance Measurement

Press the 

R

ISO

/250V/500V/1000V

 key and read the measured 

value from the display.

Example

Note

If the minimum permissible limit value is fallen short of, 
< 1 M

 or < 0.5 M

 appears for R

ISO

.

9.6

Ground Fault Test

If the test instrument recognizes an inadmissible, low insulation 
value during 

I

SC

/

R

ISO 

measurement, the instrument tests for pos-

sible ground fault.
If a ground fault occurs within the system, the error can be nar-
rowed down on the basis of the voltage ratio.

Warning!

Disconnect all solar modules from the system before 
performing the measurement.

Connect the 

PROFITEST PV

SUN

 to the module with the help of 

suitable solar plugs.

Connect the ground terminal at the 

PROFITEST PV

SUN

 to the 

module frame.

Switch the 

PROFITEST PV

SUN

 test instrument on.

Momentary open circuit voltage is displayed.
At the same time, a second voltage (

Ue

) appears in the event of a 

ground fault.
Example 

System Voltage (U0 

x 1.25)

Test Voltage

Limit Value

< 120 V

250 V

0.5 M

120 to 500 V

500 V

1 M

> 500 V

1000 V

1 M

U

0

: 0 V

I

SC

: 0.0 A

R

ISO

: < 1 M

U

0

: 0 V

I

SC

: 0.0 A

R

ISO

: ---

Uiso = 1000 V

U

0

: 0 V

I

SC

: 0.0 A

R

ISO

: > 20 M

U

0

: 144 V

I

SC

: 0.0 A

Ue: 48 V

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for PROFITEST PVSUN

Page 1: ...Operating Instructions PROFITEST PVSUN and PVSUN MEMO Test Instruments for Testing PV Modules and Strings per DIN EN 62446 VDE 0126 23 3 349 672 15 11 3 16 www com information itm com 1 800 561 8187 ...

Page 2: ...g instructions 1 USB interface cable PROFITEST PVSUN MEMO only 2 Significance of Symbols Meanings of Symbols on the Instrument Warning concerning a point of danger attention observe documentation Protection class II device continuous doubled or reinforced insulation This device may not be disposed of with the trash Fur EC mark of conformity Meanings of Symbols in the Operating Instructions Caution...

Page 3: ...sured the instrument must be removed from service and secured against unintentional use Safety of the user is no longer assured if the instrument Demonstrates apparent damage No longer performs the desired measurements Has been stored for too long under unfavorable conditions Has been subjected to mechanical stressing during transport The instrument may not be opened dismantled or modified in any ...

Page 4: ...terminal must also be tested Warning Disconnect all solar modules from the system before performing the measurement Compensation for Measurement Cable Resistance Ohmic measurement cable resistance can be subtracted from the measurement results automatically Proceed as follows in order to determine the resistance offset value Switch the PROFITEST PVSUN test instrument on Connect the measurement cab...

Page 5: ...r module under test from the system before performing measurement Attention Insulation measurement per EN 62446 is executed be tween short circuited and terminals Connect the PROFITEST PVSUN test instrument to the module with the help of suitable solar plugs Connect the ground terminal at the PROFITEST PVSUN to the module frame Switch the PROFITEST PVSUN test instrument on The module s open circui...

Page 6: ...ressing the RISO 250V 500V 1000V key The last selected value is saved after a short period of time and is then cleared from the display Example Executing the Insulation Resistance Measurement Press the RISO 250V 500V 1000V key and read the measured value from the display Example Note If the minimum permissible limit value is fallen short of 1 M or 0 5 M appears for RISO 9 6 Ground Fault Test If th...

Page 7: ...ch between objects and strings Symbol indicates whether objects or strings are active e g OBJ Object1 Standard measurement display without memory function The storage function is not activated before switch over to storage display see screenshot below Standard measurement display with activated storage function If a measurement is per formed the measured values are stored under the object and stri...

Page 8: ...lect directory C Programme PVSUN MEMO Control Center USB Driver unless you have changed the installation path Ignore the message regarding unknown provenance of the software Complete the driver installation Please check PVSUN MEMO Memory COMxx should be displayed in the device manager under Connections Start the installed program PVSUN MEMO Control Center Click on icon Search device 11 4 Creation ...

Page 9: ...e 0 to 10 Test current 200 mA Resolution 0 1 Accuracy 1 rdg 1 d No of measurements Approx 500 low resistance measurements batteries 1 5 V per IEC LR6 Display LCD Backlit multiple display Dot matrix 128 x 64 pixels Reference Conditions Ambient temperature 23 C 2 K Relative humidity 40 to 75 Battery voltage 6 V 1 V Ambient Conditions Operating temperature 0 to 40 C Storage temperature 10 C to 60 C R...

Reviews: