2

GMC-I Messtechnik GmbH

Contents

Page 

1

Scope of Delivery  ............................................................. 2

2

Significance of Symbols  .................................................. 2

3

Introduction ...................................................................... 3

4

Transport and Storage  ..................................................... 3

5

Safety Precautions ........................................................... 3

6

Use for Intended Purpose ................................................. 3

7

Operating and Display Elements   ..................................... 3

8

Initial Start-Up .................................................................. 4

9

Measurements ................................................................. 4

9.1

Testing Protective Conductor Continuity 
per EN 62446 (VDE 0126-23) (> 200 mA) ......................................4

9.2

Polarity Test / Measuring Open Circuit Voltage U

0

 

per EN 62446 (VDE 0126-23)  ........................................................4

9.3

Measuring Short-Circuit Current I

SC

 per EN 62446  ..........................5

9.4

Testing Insulation Resistance per EN 62446 ....................................5

9.5

General Insulation Resistance Measurement ....................................6

9.6

Ground Fault Test  ..........................................................................6

10

Data Storage with PROFITEST PVsun memo .................... 7

10.1

Operating and Display Elements  .....................................................7

10.2

Function Description  ......................................................................7

11

Application of PC Software  .............................................. 8

11.1

System Requirements  ....................................................................8

11.2

Installation ....................................................................................8

11.3

Initial Startup of the Device .............................................................8

11.4

Creation of Structures and Transfer to the Test Instrument  ...............8

11.5

Reading Out and Exporting Test Data  ..............................................8

12

Measurement of irradiation, temperature and inclination 
with PROFITEST 
PVsun-sor (optional)  .............................. 9

13

Characteristic Values  ....................................................... 9

14

Maintenance .................................................................. 10

14.1

Battery Replacement ....................................................................10

14.2

Housing ......................................................................................10

15

Repair and Replacement Parts Service 
Calibration Center* and Rental Instrument Service  .............10

16

Product Support  ............................................................. 10

1

Scope of Delivery

1

PROFITEST PV

SUN

 or 

PROFITEST PV

SUN MEMO

4

Batteries, 1.5 V IEC LR6 (AA)

3

Safety measurement cables, 1.5 m, red, blue and yellow: 
banana plug – banana plug

1

Solar plug adapter, red: MC3 socket – banana socket

1

Solar plug adapter, red: MC4 socket – banana socket

1

Solar plug adapter, blue: MC3 plug – banana socket

1

Solar plug adapter, blue: MC4 socket – banana socket

1

Plug-on safety test probe with socket, red

1

Plug-on safety alligator clip with socket, yellow-gray

1

Carrying case with foam insert

1

Set of operating instructions

1

USB interface cable (

PROFITEST PV

SUN MEMO

 only)

2

Significance of Symbols

Meanings of Symbols on the Instrument

Warning concerning a point of danger 
(attention: observe documentation!)

Protection class II device 
(continuous, doubled or reinforced insulation)

This device may not be disposed of with the trash. Fur-

EC mark of conformity

Meanings of Symbols in the Operating Instructions

Caution! Dangerous Voltage

Attention!

!

While performing measurements in the presence of dan-
gerous voltages, the 

PROFITEST PV

SUN

/

PROFITEST PV

SUN MEMO

 should only be operated with PPE 

safety gloves as it involves the risk of hazardous contact.

Warning!

Warning concerning a source of danger

Attention!

!

Non-observance may result in distorted measurement 
results.

Note

Please observe!

Opening of Equipment / Repair

The equipment may be opened only by authorized service per-
sonnel to ensure the safe and correct operation of the equipment 
and to keep the warranty valid.
Even original spare parts may be installed only by authorized ser-
vice personnel.
In case the equipment was opened by unauthorized personnel, 
no warranty regarding personal safety, measurement accuracy, 
conformity with applicable safety measures or any consequential 
damage is granted by the manufacturer.

Measuring Categories and their Significance per IEC 61010-1 

CAT

Definition

0

Measurements in other electrical circuits
which are not directly connected to the mains: 

e.g. electrical systems in motor vehicles and aircraft, batteries etc.

II

Measurements in electrical circuits 
which are electrically connected to the low-voltage mains: 

via plug, e.g. in household, office and laboratory applications

III

Measurements in building installations:

Stationary consumers, distributor terminals, devices connected 
permanently to the distributor

IV

Measurements at power sources for low-voltage installations:

meters, mains terminals, primary overvoltage protection devices

!

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for PROFITEST PVSUN

Page 1: ...Operating Instructions PROFITEST PVSUN and PVSUN MEMO Test Instruments for Testing PV Modules and Strings per DIN EN 62446 VDE 0126 23 3 349 672 15 11 3 16 www com information itm com 1 800 561 8187 ...

Page 2: ...g instructions 1 USB interface cable PROFITEST PVSUN MEMO only 2 Significance of Symbols Meanings of Symbols on the Instrument Warning concerning a point of danger attention observe documentation Protection class II device continuous doubled or reinforced insulation This device may not be disposed of with the trash Fur EC mark of conformity Meanings of Symbols in the Operating Instructions Caution...

Page 3: ...sured the instrument must be removed from service and secured against unintentional use Safety of the user is no longer assured if the instrument Demonstrates apparent damage No longer performs the desired measurements Has been stored for too long under unfavorable conditions Has been subjected to mechanical stressing during transport The instrument may not be opened dismantled or modified in any ...

Page 4: ...terminal must also be tested Warning Disconnect all solar modules from the system before performing the measurement Compensation for Measurement Cable Resistance Ohmic measurement cable resistance can be subtracted from the measurement results automatically Proceed as follows in order to determine the resistance offset value Switch the PROFITEST PVSUN test instrument on Connect the measurement cab...

Page 5: ...r module under test from the system before performing measurement Attention Insulation measurement per EN 62446 is executed be tween short circuited and terminals Connect the PROFITEST PVSUN test instrument to the module with the help of suitable solar plugs Connect the ground terminal at the PROFITEST PVSUN to the module frame Switch the PROFITEST PVSUN test instrument on The module s open circui...

Page 6: ...ressing the RISO 250V 500V 1000V key The last selected value is saved after a short period of time and is then cleared from the display Example Executing the Insulation Resistance Measurement Press the RISO 250V 500V 1000V key and read the measured value from the display Example Note If the minimum permissible limit value is fallen short of 1 M or 0 5 M appears for RISO 9 6 Ground Fault Test If th...

Page 7: ...ch between objects and strings Symbol indicates whether objects or strings are active e g OBJ Object1 Standard measurement display without memory function The storage function is not activated before switch over to storage display see screenshot below Standard measurement display with activated storage function If a measurement is per formed the measured values are stored under the object and stri...

Page 8: ...lect directory C Programme PVSUN MEMO Control Center USB Driver unless you have changed the installation path Ignore the message regarding unknown provenance of the software Complete the driver installation Please check PVSUN MEMO Memory COMxx should be displayed in the device manager under Connections Start the installed program PVSUN MEMO Control Center Click on icon Search device 11 4 Creation ...

Page 9: ...e 0 to 10 Test current 200 mA Resolution 0 1 Accuracy 1 rdg 1 d No of measurements Approx 500 low resistance measurements batteries 1 5 V per IEC LR6 Display LCD Backlit multiple display Dot matrix 128 x 64 pixels Reference Conditions Ambient temperature 23 C 2 K Relative humidity 40 to 75 Battery voltage 6 V 1 V Ambient Conditions Operating temperature 0 to 40 C Storage temperature 10 C to 60 C R...