GMC-I Messtechnik GmbH
65
Maintenance – Calibration
9.6
Recalibration
The respective measuring task and the stress to which your mea-
suring instrument is subjected affect the ageing of the compo-
nents and may result in deviations from the guaranteed accuracy.
If high measuring accuracy is required and the instrument is fre-
quently used in field applications, combined with transport stress
and great temperature fluctuations, we recommend a relatively
short calibration interval of 1 year. If your measuring instrument is
mainly used in the laboratory and indoors without being exposed
to any major climatic or mechanical stress, a calibration interval of
2-3 years is usually sufficient.
During recalibration* in an accredited calibration laboratory
(DIN EN ISO/IEC 17025) the deviations of your instrument in rela-
tion to traceable standards are measured and documented. The
deviations determined in the process are used for correction of
the readings during subsequent application.
By having your measuring instrument calibrated regularly, you ful-
fill the requirements of a quality management system per
DIN EN ISO 9001.
* Verification of specifications or adjustment services are not part of the
calibration. For products from our factory, however, any necessary ad-
justment is frequently performed and the observance of the relevant
specification is confirmed.
9.7
Manufacturer’s Guarantee
All
METRA HIT
digital multimeters and calibration instruments are
guaranteed for a period of 3 years after date of shipment. The
manufacturer’s guarantee covers materials and workmanship.
Damages resulting from use for any other than the intended pur-
pose or operating errors, as well as any and all consequential
damages, are excluded.
The calibration certificate confirms that the product conformed to
the specified technical data at the time of calibration. We guaran-
tee the observance of the specified technical data within the ad-
missible tolerance limits for a period of 12 months from delivery.