GMC-I Messtechnik GmbH
43
V/Hz,
, Temperature,
and A/Hz Measurements
5.7.4 Alternating Current Measurement with Clip-On Current Sensor
A AC and Hz (METRAHIT AM TECH / T
ECH
and METRAHITAM BASE /
B
ASE
only)
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter
(
METRAHIT AM TECH
/
T
ECH
: V input,
METRAHITAM BASE
/
B
ASE
: V
in-
put), all current displays appear with the correct value in accord-
ance with the selected transformation factor. The only prerequisite is
that the current sensor is equipped with at least one of the below listed
transformation factors, and that the factor has been previously selected
in the following menu (
Cl
ip
0FF
), see also chapter 6.4.
Current Clip Setup Menu
1nFo
...
SET
rAtE
...
CLIP
1
/
10
/
100
/
1000
/
0ff
Maximum allowable operating voltage is equal to the current
transformer’s nominal voltage. When reading the measured value,
additional error resulting from the clip-on current sensor must also
be taken into consideration.
(default setting:
METRAHIT AM TECH / T
ECH
:
OFF
,
METRAHITAM BASE / B
ASE
:
1:1000
)
Trans. Factors
CL
IP
Measuring Ranges DMM
Clip Types
100 mV
1 V
10 V
1:1
1 mV / 1 mA
100.00 mA
1.0000 A
10.000 A
WZ12C
1:10
1m V / 10 mA
1.0000 A
10.000 A
100.00 A
WZ12B, Z201A/B,
METRAFLEX
1:100
1m V / 100 mA
10.000 A
100.00 A
1000.0 A
Z202A/B,
METRAFLEX
1:1000
1 mV / 1 A
100.00 A
1000.0 A
(10000.0 A)
Z202A/B, Z203A/B,
WZ12C,
METRAFLEX
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
A~
Hz
003
.
50
A
AC
TRMS
A~
Hz
1:1000
050
.
10
Hz
AC
1:1000
FUNC
ENTER
COM
V
Temp
V
~
R
x
Hz
Current
R
i
R
i
~ 1 M
METRA
HIT AM BASE /
B
ASE
COM
V
Temp
A
~
R
x
Current
R
i
R
i
= 1 M
9 M
METRA
HIT AM TECH /
T
ECH
mA
METRA
HIT AM BASE /
B
ASE
METRA
HIT AM TECH /
T
ECH
V~
Hz
1M
1kHz
V~
R
i
= 1 M
R
i
= 9 M