Pulsed Transient Operation Example
This example selects CV pulsed transient operation, selects the bus as the trigger source, and programs a pulse width of 1 millisecond.
1
"INPUT OFF"
Turns off the load‘s input
2
"MODE CV"
Selects the CV mode
3
"TRIG:SOUR BUS"
Selects the bus as the trigger source
4
"VOLT: LLEV 0.5"
Sets the transient low level to 0.5 V
5
"VOLT:HIGH 1"
Sets the transient high level to 1 V
6
"TRAN:HTIM 1ms"
Sets transient high level time to 1 ms
7
"TRAN:RTIM 10us"
Sets the time for transient rising edge to 10 μs
8
"TRAN:FTIM 10us"
Sets the time for transient falling edge to 10 μs
9
"TRAN:MODE PULSE "
Selects the pulsed transient operation
10
"TRAN ON"
Activates the transient operation
11
"INPUT ON"
Turns on the load` s input
12
"*TRG"
*TRG command generates a 1 millisecond high-level pulse at the load‘s input