GEOHM 5
Instrument description
13
Step 2
Set the function switch to
R
EARTH
position, the following menu will be displayed:
Figure 7. Earth Resistance initial menu
Step 3
Press START key and view the test result. (If you want to perform
more than one measurement press the
START
key and keep it
pressed until the result has stabilized, then release the key. The last
result will stay displayed.)
Press the DISP key to check the resistances of both the potential
(
Rp)
, and the current (
Rc)
test probes. After a short period
the main
result will automatically be displayed again.
Store the displayed result for documentation purposes, see instructions on how to
use memory in chapter 3.1.1. Storing of test results.
Notes!
When an external voltage higher than 20 Vac/dc. is present between H and E or
ES and S test terminals, the Earth Resistance measurement will not be carried out
after pressing
START
key, but the message
> 20 V
will be displayed, marked with
!
symbol!
If a noise voltage higher than approx. 5 V is present between the H and E or ES and
S test terminals, a cursor mark pointing to the
“
” (noise) symbol will be
displayed, indicating that the test result may not be correct!
If the resistance of the current or potential rod is too high (>(4 k
+ 100 R
E
) or
>50 k
, whichever is lower), then the test result will be marked with
symbol and
the cursor will appear on
Rc
and / or
Rp
.
If the test result is out of measurement range (open test leads),
>19,99 k
message will be displayed!