GoodLife™
About the System
Intended Use
Principle of the Test
GoodLife Blood Glucose Monitoring System(Cloudia)
is intended to be used for the quantitative measure-
ment of glucose (sugar) in fresh capillary whole
blood sample drawn from the fingertips only. It is
intended to be used for single patient and should not
be shared. It is not indicated for the diagnosis or
screening of diabetes or for neonatal use.
The GoodLife System is an electrochemical biosensor
system that measures the amount of electric current
produced and displays the result as a blood glucose
level.
The GoodLife Blood Glucose Monitoring System
(Cloudia) (GoodLife System) is designed to provide
blood glucose measurements with easy and comfort-
able testing. The system requires 0.5μL of blood
sample and 5 seconds for the test to complete.
The GoodLife System consists of:
GoodLife Blood Glucose Meter (Cloudia), GoodLife
Blood Glucose Test Strips (Cloudia), Check Strip,
GoodLife Glucose Control Solution(Level I & II), Safety
Lancet (optional for
purchase).
These products are intended to be used together to
get accurate blood glucose test results. Do not use
other test strips or control solutions with your meter.
NOTE:
Not for use on patients who are dehydrated, hypo-
tensive, in shock, or in a hyperosmolar state.ÿ