2.TECHNICAL SPECIFICATIONS
Sensitivity
2mV~5V/DIV, 11 steps in 1-2-5 sequence
Sensitivity Accuracy
±
3% (5 DIV at the center display )
Vernier Vertical Sensitivity
Continuously variable to 1/2.5 or less than
panel-indicated value
Frequency Bandwidth(-3dB) DC ~ 200MHz (5mV/DIV:DC~150MHz)
(2mV/DIV:DC ~ 20MHz)
Rise Time
1.75ns(5mV/DIV:2.33ns) (2mV/DIV:17.5ns)
Signal Delay
Leading edge can be monitored
Maximum Input Voltage
400V (DC+AC) at 1kHz or less
Input Coupling
AC, DC, GND
Input Impedance
1M
Ω
±2% // 25pF approx.
Vertical Modes
CH1,CH2,DUAL(CHOP/ALT),ADD(DIFF mode can
Be established when the CH2 is in the INV mode)
CHOP Repetition Frequency Approx. 250kHz.
Polarity (INV)
CH2 only
Bandwidth Limited
20MHz
VERTICAL
DEFLECTION
SYSTEM
Common-mode
Rejection Ration
50:1 or better at 50kHz
Trigger Modes
AUTO, NORM, TV
Trigger Source
CH1, CH2, LINE, EXT. EXT/10
Trigger Coupling
AC, DC, HFR, LFR, NR
Trigger Slope
+/- polarity or TV sync polarity
Mode
Frequency INT
EXT
10Hz~20MHz 0.35DIV
50mVpp
AUTO
20MHz~200MHz
1.5DIV
150mVpp
DC~20MHz 0.35DIV
50mVpp
NORM
20MHz~200MHz
1.5DIV
150mVpp
Trigger Sensitivity
TV Sync
signal
1DIV
200mVpp
Trigger Level Range
INT : ± 4 DIV or more
EXT : ± 0.4 V or more
EXT/10 : ± 4 V or more
TV-V, TV-H, TV-L
TV-Line Selection
Standard
Field 1
Field 2
NTSC(525H) 1H~263H
1H~262H
PAL (625H)
TV Sync
SECAM(625H)
1H~313H 1H~312H
Max. External Input Voltage 400V (DC + AC peak) at 1kHz
TRIGGER
SYSTEM
External Input Impedance
1M
Ω
± 5%
∕
∕
25pF approx.
GOS-6200 OSCILLOSCOPE
USER MANUAL
⎯
5
⎯
Horizontal Modes
MAIN(A), ALT, DELAY(B)
A (main) Sweep Time
20ns~0.5sDIV, continuously variable
(UNCAL)
B (delay) Sweep Time
20ns~50msDIV
Accuracy
± 3% (± 5% at × 10 MAG)
Sweep Magnification
× 10 (maximum sweep time 2nS/DIV)
Hold Off time
Variable
Delay Time
1us~5s
Delay Jitter
1/20000 or less
HORIZONTAL
DEFLECTION
SYSTEM
Alternate Separation
Variable
Sensitivity Accuracy
X-axis, Y-axis selectable
X-axis: CH1, CH2
Æ
2mV~5V/DIV ± 3%
EXT
Æ
0.1V/DIV ± 5%
EXT/10
Æ
1V/DIV ± 5%
Y-axis: CH1,CH2
Æ
2mV~5V/DIV ± 3%
X-axis Bandwidth
DC~500kHz (-3dB)
X-Y
OPERATION
Phase Error
3
°
or less at DC~50kHz
Cursor Measurement Function
Δ
V,
Δ
V%,
Δ
VdB,
Δ
T,1/
Δ
T,
Δ
T%
,
Δ
Θ
.
Cursor Resolution
1/100 DIV
Effective Cursor Range
Vertical: ± 3 DIV; horizontal: ± 4 DIV
CURSOR
READOUT
FUNCTION
Panel setting
Vertical: V/DIV (CH1
,
CH2),UNCAL, ADD,
INV, P10, AC/DC/GND.
Horizontal: S/DIV (MTB, DTB), UNCAL
x
10MAG, Delay time, Hold-off.
Trigger: Source, Coupling, Slope, Level ,
TV-V/TV-H/TV-L.
Others: X-Y, LOCK, SAVE/RECALL
MEM 0-9.
Parameter Function
FREQ, PERIOD,
±
WIDTH,
±
DUTY,
(+ or – polarity selected by trigger slop.)
Display Digits
Max. 6-digits, decimal.
Frequency Range
50Hz~200MHz.
Accuracy
1kHz~200MHz:
±
0.01%
50Hz~1kHz :
±
0.05%
AUTO
MEASUREMENT
FUNCTION
Measuring Sensitivity
>2div(measuring source selected from CH1
and CH2 as synchronous signal source.)