Be Smart, Be Compact!
i) Drag the display from bottom to up as follow image by finger. To back drag the display to up again.
0.0
20
2) Calibration jig
How to use
6
ii) Connect the calibration jig and click Calibration icon.
2.0
0.5
1.0
0.0
0.0
0.0
0.0
* Accuracy check
Note
If the figure is not 2.9. Need to calibrate.
- Connect the calibration jig on the probe jack. And turn on the device.
- Check the measured figure shows 2.9.
* Calibration
- If the figure doesn't show 2.9 from the accurate check, need to calibrate as follow step:
iv) Wait till “CALIBRATION SUCCESS” message.
And if "CALIBRATION SUCCESS" message is shown,
calibration is done.
iii) click “YES” button.
2.0
0.5
1.0
0.0
0.0
2.9
CALIBRATION?
CALIBRATION
SUCCESS
CONFIRM
Note
If "CALIBRATION
SUCCESS FAIL"
message is shown,
try calibration again.
CALIBRATION
FAIL
CONFIRM
DRAG