SmartBox
®
1 / SmartBox
®
2 / SmartBox
®
3
12 / 16
Part No. 28 100 57 e
Menu
Setting
Description
13.Rounding Automatically
Without rounding
20L
50L
100L
200L
500L
1.000L
Default settings
minimal increments
Rounding increments in relation to the set volume
set with [+] / [-] keys
14.–17.Exit
Press [Enter] to return to display mode
18.Language Language:
German, English, French
[+] / [-] / [Enter]
Names:
ESC
[+] / [-] / [Enter]
Name Tank 1
: Suggested name
Letters can be changed with [+] / [-] / [Enter]
19.Exit
Press [Enter] to return to display mode
20.LCD
display
Contrast: 24
Set the contrast of the LCD display
21.Device
info
Displays information about
Software version: V4.00
(e. g.)
Serial number: Tank 1: SN=1234
(e. g.)
Gain: X0 = 4.05 mA; B =
1268
(Tank 1)
22.Test
current
Test function for the current mA value of the
measurement probe:
ADC = 28A0 = 04.00 mA
If level probe is not submerged, the value should
be close to 4 mA. Tolerance range is 3.7...4.3
mA. For larger deviations, see menu item 9.
23.Test
relay
SBox 2+3
Rel1 ON / OFF
Relay test
24.Reset
ESC
Reset the device software:
Exit this function without executing it.
Restart
Initialisation. The device software restarts and
keeps all device settings.
Factory settings
Complete reset of all parameters to the original
delivery status.
26.Exit
Press [Enter] to return to display mode