2. USER’S MANUAL FOR OPERATION
2.4.1 - You can not connect to the GNOM and other devices (accessory) to the
mains with the following problems:
• Wrong - incorrect wire connections L and N conductors in the appropriate slot.
• without protective earthing conductor.
Note. Checking the correct wiring connections (phase conductor - L) inspects the
operator. (E.g. by phase tester)
2.4.2 Checking the RCD
Carry out before any underwater operation at the new location as follows:
• If the requirements of paragraph 2.4.1 are met and connect to the power supply
only power module
• Switch on mains voltage button on the control panel
• After 5 seconds press the "TEST" button on the residual current device. It would
immediately lead to a safe shutdown of module
• Switch off the mains voltage button on the control panel. Switch on the current
circuit breaker
• Again switch on the power supply button on the control panel. Residual current
device must remain switched on.