IDENTIFICATION OF EQUIPMENT

Description of the Equipment

Optilight Color is the latest generation of LED photo activation devices.

This abbreviation is the acronym for Light Emitting Diode, a totally different way of emitting light,

in comparison to the halogen conventional devices. Unlike traditional equipments, which generate

light in wide wave spectrum and a lot of heat, this technology allows the emission of cold light in

a precise wavelength for the activation of several dental products apply.

LED technology, recently introduced in odontology, has brought several advantages to

photopolymerizers on direct restoration in composed resin. Besides infinitely more durable, LEDs 

have become the most compact, ergonomic and easy-to-install and transport equipments. The 

cold light emission and in precise wavelengths assure a safe polymerization of activated composite

by the camphorquinone, without the risk of tooth warming, pulp injuries or discomfort for both 

operator and patients.

The safety and efficiency of the LEDs, now with high energy emission, are available to all the clinical

procedures that require light power for photo activation.

The wavelength from 420nm to 500nm associated with the high energy emitted by Optilight Color

enables the multi functionality of this device:

• Direct restoring processes: composite resins, ionomers and adhesives.

• Indirect restoring: laminated adhesive cementation, inlays, aesthetic pins and metalfree diadems.

• Activation of material photo active as sealants, surgical cements and the basis of a foraging.

Designed and built within the most advanced technology in order to provide results within the 

specifications set by the highest world odontology authorities.

Equipped with a power supply with an automatic bi-volt key that allows to use the equipment in 

any supply voltage between 100 to 240V ~ - 50/60Hz.

Digital control on the display in the handpiece itself.

Variation of choosing the operation time (5,10,15 and 20 seconds).

It has 3 application modes: Continuous, Ramp and Pulse:

• 

Continuous

: Maximum mode and continuous light intensity (same luminosity from the beginning

to the end of the polymerization). 

• 

Ramp: 

Gradual light intensity mode, increases gradually.

Physical Principle used by the Equipment

The physical principle is the emission of a light for polymerizing photo sensible substances, since 

the equipment is featured with a cold light emitter (LED) with the wavelength comprised within 

420nm a 500nm (blue light), which has the ideal intensity for integrating with camphorquinone.

Indication of Equipment

This equipment is for dental use only aiming to photo polymerize photo sensible substances through
the emission of a blue light. 

It is used in various dental procedures such as restorative procedures, bracket bonding and activation 

of photoactivated materials such as sealants, dental bases, for exclusive dental use, and must be 

used and handled by a qualified person (duly regulated professional, according to local law of the 

country). following the instructions in this manual.

Contraindications

Systemic diseases (canc er, cardiovascular diseases serious diseases, the blood system, the immune 

system The disease, ...).

Ongoing and topical treatment of certain systems (anticoagulant therapy, chemotherapy, 

radiotherapy, ...).

Ask patients if they have a cardiac pacemaker or other system implanted before you Intended use

start the treatment.

Summary of Contents for Optilight Max

Page 1: ......

Page 2: ...979 736 0001 45 Insc Est 582 002 897 114 Rod Abrão Assed Km 53 450m Cx Postal 782 CEP 14097 500 Ribeirão Preto S P Brasil Telefone 55 16 3512 1212 Technical Duties Daniel R de Camargo CREA SP 5062199650 Registration ANVISA 10069210078 For improved safety Read and understand all the instructions contained in the Instructions for use before installation or operation of this equipment The Instruction...

Page 3: ...s during installation of the equipment 20 Precautions and warnings during the use of the equipment 20 Precautions and warnings after the use of equipment 21 Precautions and warnings during cleaning and disinfection of the equipment 21 Precautions in case of alteration in the functioning of equipment 21 Precautions for the Reduction of the Environment Impact 21 Precautions to be adopted against for...

Page 4: ...details that might orientate you in its correct use as well as in the solution of small problems that might effectively occur We advise its complete reading and conservation for further consults IDENTIFICATION Technical Name Equipment for resins photopolymerizers Trade Name Curing Light Optilight Color Model Optilight Color ...

Page 5: ...nced technology in order to provide results within the specifications set by the highest world odontology authorities Equipped with a power supply with an automatic bi volt key that allows to use the equipment in any supply voltage between 100 to 240V 50 60Hz Digital control on the display in the handpiece itself Variation of choosing the operation time 5 10 15 and 20 seconds It has 3 application ...

Page 6: ...a forced ventilation system therefore avoiding noise emission We observe that the light emitted by the Optilight Color is completely contained within the absorption interval of the photo initiator and is therefore 100 used while conventional devices which use halogen lamps have a big part not used in the process Optilight Color does not produce heating because uses semiconductors LEDs as light emi...

Page 7: ... conductor 02 Handpiece 03 Power supply unit 04 Base A Commands B Indicative LED for the battery level A Time adjust button Visor display Continuous mode Ramp mode Pulse mode On Off button Application mode selector button B Green Full charge Red Loading LED NOTE Available in colors ...

Page 8: ...tration Therefore when buying the product verify the technical compatibility between the equipment coupling and accessories The use of any part accessory or material neither specified nor foreseen in those use operation instructions and it is of the user s entire responsibility BASE OCULAR PROTECTOR MODULES ACCESSORIES OPTIONS AND MATERIALS OF CON SUMPTION BATTERY SENT INSIDE EQUIPMENT ...

Page 9: ...ries with charge time of at least 8h Light conductor Optic fiber 100 coherent that ensures the light passage without losses ø8mm 60º curve Timer 5 10 15 and 20 seconds Time Sounder A beep in every 05 seconds Triggering Through the handpiece s button Handpiece s body Injected in ABS Net weight 0 389 kg Gross weight 0 640 kg Protection against the electrical shock Class II Equipment Applicated part ...

Page 10: ...r its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions ABNT NBR IEC CISPR 11 Class B The Optilight Color is suitable for use in all establishments including domestic establishments and those establishments directly connected to the public low voltage power supply network with specific requirement Emis...

Page 11: ... 5 kV 1 kV modo diferencial 0 5 kV 1 kV 2 kV modo comum M a i n s p o w e r q u a l i t y should be that of a typical commercial or hospital environment Reduction interruption and variance of voltage in power supply input lines IEC 61000 4 11 100 Ut 100 dip in Ur for 0 5 cycle 100 Ut 100 dip in Ut for 1 cycle 30 Ut 70 dip in Ut for 25 30 cycles 100 Ut 100 dip in Ut for 250 300 cycles 100 Ut 100 di...

Page 12: ...rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and ...

Page 13: ...ration distance according to frequency of transmitter 150 kHz to 80 MHz d 1 2 p 80 Mhz to 800 MHZ d 1 2 p 80 MHz to 800 Mhz d 2 3 p 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the ...

Page 14: ...14 Dimensions mm TECHNICAL SPECIFICATIONS ...

Page 15: ...ight Fragile handle with care Temperature limitation Type B General warning Attention Recyclable Refer to the instruction manual Grounding at several points of the equipment indicates the condition of being grounded Serial number Model Number Manufacturing date Manufacturer 12 TECHNICAL SPECIFICATIONS ...

Page 16: ...k and feel that it was correctly embedded 3 Insert the ocular protector on the light conductor Charge the battery during 08 hours before using the equipment for the first time 1 2 3 4 Place the base on an stable site connect the power supply unit s cable to the handpiece and immediately connect the power supply unit to an electrical socket 5 Let the battery charging for at least 4 hours 4horas 4 5...

Page 17: ...emove the light conductor protection cover and take the handpiece to the patient s mouth and position the light conductor at a safe distance fig 4 For starting the polymerization cycle press the trigger Just trigger again to interrupt fig 5 1 2 3 Press Press Press 4 5 10mm beep beep beep ATTENTION Low battery When the display shows the symbology and sound the beep indicating low battery please rec...

Page 18: ...mW cm Tolerance of 200 mW cm IMPORTANT Never address the light beam to the eyes Protect the visual field using the ocular protector The ocular protector has the goal of filtrate only the blue light that operates in the photo polymerization of resins for protecting the sight and still allows that the room light has a path to the operative field After using it keep the light conductor protected by t...

Page 19: ... Environmental operation condition Room temperature range for functioning 10ºC to 40ºC Recommended room temperature range 21ºC to 26ºC Relative humidity range for functioning 30 to 75 Atmospheric pressure range 700 hPa to 1060 hPa 525 mmHg to 795 mmHg Operation altitude 2000m ATTENTION The Equipment maintains its condition of safety and efficacy provided that it is maintained stored as mentioned i...

Page 20: ...ions such as vacations clean and disinfect the equipment This equipment is not sensible to electrical interference electrostatics and pressure since cleaning items maintenance transport and operation of this manual are observed However an electromagnetic environment can interfere in its normal operation The equipment must only be operated by technicians duly qualified and trained Dental Surgeons a...

Page 21: ...he conductor must be clean and sterilized at 134ºC before being used in the next patient Before cleaning the equipment disconnect it from the electricity Avoid spilling water or other liquids inside the equipment what might cause short circuits Do not use micro abrasive material or scouring pad in cleaning do not use organic solvents or detergents that can contain solvents like ether stain remover...

Page 22: ...eding to be cleaned and disinfected Attention Do not use chlorine containing disinfectants and do not use ultrasonic cleaning Disinfection The Handpiece and battery can be surface wiped with 75 ethanol Light conductor and ocular protector can be immersed in 75 ethanol for 30 minutes Drying After cleaning and disinfection please dry it is recommended to use compressed air to dry Warning 1 Handpiece...

Page 23: ...is related to some of the listed items under the item Unpredictable situation cause and solution If it s not possible to solve the problem shutdown the equipment and demand the presence of a Alliage technician from the nearest resale or ask through the Attendance Service Alliage 55 16 3512 1212 Corrective Maintenance The supplying of the circuits diagram Part lists or any other information that pe...

Page 24: ...attery failure Restore the connection Get in touch with Alliage technical assistance The equipment is not polymerizing the resins Non appropriated resin for the wavelength of LED photopolymerizers Light conductor incorrectly Resin residue in the light conductor Light conduct or with protection cover Buy the appropriated resin for the wavelength of the polymerizer that is which contains photo initi...

Page 25: ...presentative We call your attention to our authorized resellers chain Only this chain will keep your equipment constantly new because it has trained technical assistant and specific tools for the correct maintenance of your device Whenever you need demand the presence of a Alliage technician from the nearest resale or ask through the Attendance Service Alliage 55 16 3512 1212 EQUIPMENT S WARRANTY ...

Page 26: ...26 ...

Page 27: ...27 ...

Page 28: ......