24
Attention
Upon coming across any problem in operation, follow the instructions below to check and
repair the problem, and/or get in touch with your representative.
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
Imprevistos
Provável Causa
Soluções
Problem
Probable cause
Solution
-Equipment inoperative.
“Display with no show”.
-Handpiece battery without
charge.
-Overheating Protection.
-activated “Error code: Er”.
-Recharge the handpiece on
the base for 4 hours.
-Wait the cooling for some
minutes.
- The battery doesn’t charge.
-Bad connection between
the supply source and the
handpiece.
-Battery failure.
-Restore the connection.
-Get in touch with Alliage’
technical assistance.
-The equipment is not
polymerizing the resins.
-Non appropriated resin
for the wavelength of LED
photopolymerizers.
-Light conductor incorrectly.
-Resin residue in the light
conductor.
- L i g h t c o n d u c t o r w i t h
protection cover.
-Buy the appropriated resin
for the wavelength of the
polymerizer, that is, which
contains photo initiators with
camphorquinone.
-Correctly attach the light
conductor fixed.
-Clean the light conductor.
-Remove the protection cover
of the light conductor.
-Inadequate Luminous power.
-Light conductor incorrectly
fixed.
-Problems with light conductor.
-Reduced battery’s capacity.
-Fix correctly the light
conductor.
-Substitute the light conductor.
-Contact Alliage’ technical
assistance.
Summary of Contents for Optilight Max
Page 1: ......
Page 14: ...14 Dimensions mm TECHNICAL SPECIFICATIONS ...
Page 26: ...26 ...
Page 27: ...27 ...
Page 28: ......