CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Sterilization
Follow the instructions of the manufacturer of the appliance.
Only sterilize clean and lubricated handpieces.
Clean, grease and remove the overflow oil in the apparatus according to the previous indicate, put
it in the surgical grade paper (not provided by GNATUS) and sterilize it in autoclave until 135°C
(275°F) through a cycle.
The bit breaker can be placed into an autoclaved up to 135°C (275°F) during one cycle.
Preventive Maintenance
The equipment must suffer routinely measurements, following the current legislation of the country.
But, never with a period superior to 3 years.
For protecting your equipment, look for a GNATUS’ technical assistance for periodic reviews as
preventive maintenances.
Corrective Maintenance
GNATUS states that the supplying of the circuits’ diagram, Part lists or any other information that
permits the technical assistance by the user, can be requested, since previously agreed between
the buyer and GNATUS.
Warning
Sterilize before 1st use, product do not provided sterilized.
Not valid or recommend to the sterilization by ethilene oxyde.
Do not sterilize in an oven.
Before sterilization, connect it to the delivery unit and activate it for a few seconds to
eliminate the excess of oil.
If the excess of oil is not eliminated, it may cause burnt residue in the bearing.
Attention
In case of the equipment presents any abnormality; check if the problem is related to some
of the listed items under the item Unpredictable (situation, cause and solution).
If it’s not possible to solve the problem, shutdown the equipment and demand the presence
of a GNATUS’ technician from the nearest resale, or ask through the Attendance Service
GNATUS: + 55 (16) 3512-1212.