English
- Check the voltage of the equipment at the moment of executing the electrical
installation.
- Position the equipment in a place where it will not get wet.
- Install the equipment in a place where it will not be damaged by pressure, temperature,
humidity, direct sunlight, dust, salts, or sulfur compounds.
- The unit should not be submitted to inclination, excessive vibrations, or blows (including
during transportation and handling).
- This equipment was not planned for use in an environment where vapors, anesthetic
mixtures infl ammable with air, or oxygen and nitrous oxide can be detected.
- Before the first use and/or after long interruptions from work such as vacations, clean
and disinfect the equipment..
Precautions and warnings “during the installation” of
quipment
PRECAUTIONS, RESTRICTIONS AND WARNINGS
This equipment must be transported and stored observing the following directions:
- Avoid falls and impacts;
- Keep it dry, do not expose it to rain, water drops or wet floor;
- Keep it away from water and direct sunlight, and in it original wrapping;
- Don’t move it over irregular surfaces, protect it from rain and observe the maximum
stack quantity specified in the packaging;
- Transportation and storage temperature range: -12°C to 50°C;
- Transportation and storage relative humidity range: 0°C to 90°C;
Transportation and storage
The Equipment maintains its condition of safety and efficacy, provided that it is
maintained (stored) as mentioned in this instruction of use. Thus, the equipment
will not lose or alter its physical and dimensional features.
- The equipment has been planned not to be sensitive to interference such as magnetic
fields, external electrical factors, electrostatic discharge, pressure or variance of pressure,
provided that the equipment is installed, maintained, clean, preserved, transported and
operated as per this instruction for use.
Sensitivity to environmental conditions in normal situations
of use
Grounding:
Grounding is compulsory. Only in this way the user is in total safety. It is only the
user’s responsibility for not grounding, implying in warranty loss.
Use three-prong plug. Never connect the grounding to the neutral.