GN CX-WLTFM User Manual Download Page 5

 

Numerical after "c" ( such as 46.33 ) is the modified flow coefficient (k) value, the number of 
flashes that can be modified, and modify the scope: 0 to 9, the position can be shifted by 
press the “CALIBRATE” button shortly 

 

To press “ DISPLAY” shortly can increase the numerical from 0 to 9, increase only , can not 
reduce. 
   
After modification, press “DISPLAY” and “CALIBRATE” buttons simultaneously, it can 
return to normal working status.   

 

User manual flow coefficient (k) calibration correction formula: 

 

When the user found that the actual traffic flowed liquid flow has a larger error with 
displayed flow, using the following formula to re-calculated k value 

 
 

       

                                     

0

0

L

k

l

k

×

=

             

Note

 

 

 

l= flow rate of flow meter     

 

k

0

= current flow

not to amend before

k value 

 

L

0

user graduated cylinder measures the value 

 Users to manually modify the k value, for example: 

 Clear "0" TTL1 value or remove the batteries, to clear TTL1, TTL2 "0", fitted with batteries, 
button is shifted to TTL1 status, start the measurement of liquid flow, the flowed liquid 
enters into the standard cylinder. Record the displayed flow value and graduated cylinder 
flow value after the liquid flows out, such as displayed flow l=95.56 LTR, graduated cylinder 
measures the value L

0

100 LTR, current k value ( not to amend before ) k

0

=40, Calculated 

as follows 

 

224

.

38

100

40

56

.

95

=

×

=

k

 

 

As it only keeps the 2 numerical after two decimal, k

38.22, to modify k value to 38.22 is 

okay 

 

Tel:86-21-57632436                                                   Fax:86-21-57634942

Shanghai Cixi instrument Co., ltd

www. gnflowmeter.com

Summary of Contents for CX-WLTFM

Page 1: ...Usermanual CX WLTFM TURBINEFLOWMETER CX WLTFM TURBINEFLOWMETER Tel 86 21 57632436 Fax 86 21 57634942 Shanghai Cixi instrument Co ltd www gnflowmeter com ...

Page 2: ...flow coefficient k and user manual calibration of the flow coefficient k calibration 2 Functions as below Two cumulate which can be reset TTL1 cumulate Clear TTL2 the total accumulation can not clear zero it can be shifted by the buttons Unit selection The selection buttons to operate the flow units liters and gallons U S Sensor pulse frequency range 5Hz 5000Hz Low power consumption Two pieces of ...

Page 3: ...ss CALIBRATE can shift the unit LTR and GAL and the corresponding characters will be showed on the LCD display To press DISPLAY and CALIBRATE at the same time about 2 seconds will enter the user manual flow coefficient k calibration status B LCD Description TTL1 current flow can be cumulative in which can be cleared zero TTL2 current flow displayed is total cumulative in which can not be cleared z...

Page 4: ...ut about 30 seconds control panel will record the flow value calculate the flow coefficient k values and deposited CPU of the permanent preservation of non volatile memory even removed the battery discharge coefficient k will not lose value If the flow of calibration error over more than the limited 10 the panel will consider the current calibration value is invalid and will not record the current...

Page 5: ...0 L k l k Note l flow rate of flow meter k0 current flow not to amend before k value L0 user graduated cylinder measures the value Users to manually modify the k value for example Clear 0 TTL1 value or remove the batteries to clear TTL1 TTL2 0 fitted with batteries button is shifted to TTL1 status start the measurement of liquid flow the flowed liquid enters into the standard cylinder Record the d...

Page 6: ...tenance Battery electrode should be inspected once a year No need to remove the instrument from the pipeline system the batteries can be replaced V Batteries replacement Remove the four angle screws from the instrument s control panel take the computerized electronic instruments LCD screen from the turbine flow Take out the batteries Check battery electrode if rust and clear it Install new batteri...

Page 7: ...ilure 1 Removed the computer and if necessary check and reinstall or replace the batteries 2 Factory contact Normal flow rate but the instrument does not count when pressing the DISPLAY button the instrument boots 1 On site calibration method is not correct 2 Impeller jammed or damaged 3 Sealant material wrapped around the impeller 4 The computer failure Re check on site or check by Factory Remove...

Reviews: