Question
Answer
Why do my hearing aids not turn on auto-
matically, although they have been in the charger
for an extended period of time?
If you leave your hearing aids in the charger for more than
24 hours, the charger will go into standby mode and turn off
the power of the hearing aids. If the charger* runs out of
battery power, it will instruct the hearing aids to turn off to
save energy. When you remove the hearing aids from the
charger, they will not turn on automatically as normally.
Press the push button for 5 seconds to turn the hearing aids
on manually.
*: Premium charger only
Does the battery heat up when charging?
There is a slight temperature increase at the end of the char-
ging process
Can I charge at low temperatures?
If the hearing aid temperature is below 0 °C (32 °F), it will not
charge immediately. Charging must take place between
0 °C (32 °F) and 40 °C (104 °F)
Can I charge at high temperatures?
The operating temperature range for the charger and the
hearing aids is 0 °C (+32 °F) to +40° C (+104 °F)