GMI D5263S Instruction Manual Download Page 2

 

D5263 

- SIL 2 Load Cell/Strain Gauge Bridge Isolating Repeater 

G.M. International ISM0227-8 

General Description: 

The single channel DIN Rail Load Cell/Strain Gauge Bridge Isolating Repeater D5263S module is a unit suitable for applications requiring SIL 2 level (according to IEC 61508:2010) in 
safety related systems for high risk industries. 
The unit acts as a transparent galvanic isolated interface installed between a weighing indicator in Safe Area and a load cell (or group of load cells) in Hazardous Area; it appears at the 
terminals of the indicator as a single load cell equivalent to the one in the field. It provides a fully floating power supply voltage with remote sensing capability to load cell located in 
Hazardous Area and repeats, while isolating, the mV signal output to drive a load in Safe Area depending on the host system reference voltage. Up to four 350 

 load cells, or five 450 

 load cells, or ten 1000 

 load cells can be connected in parallel. The Voltage reference (Safe Area side) is set as an external supply.  

Function:

 

1 channel I.S. input from strain gauge signals, provides 3 port isolation (input/output/supply) and repeats, as a transparent unit, the bridge signal output. 

Signalling LED:

 

Power supply indication (green). 

EMC:

 

Fully compliant with CE marking applicable requirements. 

 

Technical Data 

Characteristics 

Supply: 

24 Vdc nom (18 to 30 Vdc) reverse polarity protected, ripple within voltage limits 

 5 Vpp. 

Current consumption @ 24 V:

 85 mA with four 350 

 load cells connected, typical. 

Power dissipation:

 1.8 W with 24 V supply and four 350 

 load cells connected typical. 

Isolation (Test Voltage): 

I.S. In / Out 1.5 KV; I.S. In / Supply 1.5 KV; Out / Supply 500 V. 

Input:

 

up to four 350 

 load cells in parallel or up to five 450 

 load cells in parallel or up to ten 1000 

 load cells in parallel. 

A/D conversion time: 

12.5 ms 

Bridge supply voltage: 

4.0 Vdc nominal. 

Bridge output signal: 

1 to 4 mV/V. 

Line resistance compensation: 

 10 

Output:

 

Same as Input signal. 

Output impedance: 

500 

 typical. 

Excitation voltage: 

 4.0 V typical, 

 15 V maximum. Externally applied. 

Transfer characteristic: 

linear based on mV input. 

Response time:

 

 100 ms (10 to 90 % step change). 

Performance:

 

Ref. Conditions 24 V supply, 23 ± 1 °C ambient temperature. 

Calibration accuracy after system calibration: 

 ± 0.003 % of full scale of input range. 

Linearity accuracy:

 

 ± 0.002 % of full scale of input range. 

Supply voltage influence: 

 ± 0.002 % of full scale for a min to max supply change. 

Temperature influence: 

 ± 0.002 % of full scale of input range for a 1 °C change. 

Compatibility:

 

CE mark compliant, conforms to Directives:  
2014/34/EU ATEX, 2014/30/EU EMC, 2014/35/EU LVD, 2011/65/EU RoHS. 

Environmental conditions:

 

Operating:

 temperature limits –40 to + 70 °C, relative humidity max 95 % non condensing, up to 55 °C. 

Storage:

 temperature limits – 45 to + 80 °C. 

Max altitude:

 2000 m a.s.l. 

Safety Description: 

ATEX:

 II 3(1)G Ex ec [ia Ga] IIC T4 Gc, II (1)D [Ex ia Da] IIIC, I (M1) [Ex ia Ma] I 

IECEx:

 Ex ec [ia Ga] IIC T4 Gc, [Ex ia Da] IIIC, [Ex ia Ma] I 

EAC-EX: 

2Ex nA [ia Ga] IIC T4 Gc X, [Ex ia Da] IIIC, [Ex ia Ma] I 

CCC:

 Ex nA [ia Ga] IIC T4 Gc, [Ex ia Ga] IIC, [Ex iaD] 

CML:

 Ex nA [ia Ga] IIC T4 Gc, [Ex ia Da] IIIC 

associated apparatus and non-sparking electrical equipment. 

Uo/Voc = 7.2 V, Io/Isc = 177 mA, Po/Po = 471 mW at terminals 13-14-15-16-17-18. 

Um = 250 Vrms, -40 °C 

 Ta 

 70°C. 

Approvals:  

TUV 15 ATEX 170897 X conforms to EN60079-0, EN60079-7, EN60079-11. 

IECEx TUN 16.0005X conforms to IEC60079-0, IEC60079-7, IEC60079-11. 

RU C-IT.EX01.B.00018/19 conforms to GOST 31610.0,GOST 31610.11, GOST 31610.15. 

CCC 2020322316000978 conforms to GB3836.1, GB3836.4; GB3836.8, GB12476.1, GB12476.4. 

CML 20JPN2136X for CML approval. 

DNV No. TAA00001U0 and KR No.MIL20769-EL002 for maritime applications. 

SIL 2 conforms to IEC61508:2010 Ed.2. 

Mounting: 

EN/IEC60715 TH 35 DIN-Rail, with or without Power Bus. 

Weight:

 about 165 g. 

Connection:

 by polarized plug-in disconnect screw terminal blocks to accommodate terminations up to 2.5 mm

2

 (13 AWG). 

Location:

 installation in Safe Area or Zone 2, Group IIC T4. 

Protection class:

 IP 20. 

Dimensions:

 Width 22.5 mm, Depth 123 mm, Height 120 mm. 

 

 

Summary of Contents for D5263S

Page 1: ...263 SIL 2 Load Cell Strain Gauge Bridge Isolating Repeater G M International ISM0227 8 SIL 2 Load Cell Strain Gauge Bridge Isolating Repeater DIN Rail and Power Bus Model D5263S D5263S INSTRUCTION MANUAL ...

Page 2: ...esistance compensation 10 Ω Output Same as Input signal Output impedance 500 Ω typical Excitation voltage 4 0 V typical 15 V maximum Externally applied Transfer characteristic linear based on mV input Response time 100 ms 10 to 90 step change Performance Ref Conditions 24 V supply 23 1 C ambient temperature Calibration accuracy after system calibration 0 003 of full scale of input range Linearity ...

Page 3: ... to the instruments associated with the barrier 20 16 15 19 18 14 17 13 1 2 3 4 D5263 5 6 7 8 9 10 11 12 PWR SIL 2 13 14 15 16 4 3 2 1 8 7 6 5 12 11 10 9 17 18 19 20 Front Panel and Features Terminal block connections Ordering Information Model D5263S Power Bus and DIN Rail accessories Connector JDFT050 Cover and fix MCHP196 Terminal block male MOR017 Terminal block female MOR022 HAZARDOUS AREA SA...

Page 4: ...rs D5263 Associated Apparatus Parameters D5263 Terminals Function Diagram HAZARDOUS AREA ZONE 0 ZONE 20 GROUP IIC SAFE AREA ZONE 2 GROUP IIC T4 Uo Voc 7 2 V Ui Vmax 13 14 15 16 17 18 Io Isc 177 mA Ii Imax Po Po 471 mW Pi Pi Must be Hazardous Area Hazardous Locations Device Parameters D5263 Associated Apparatus Parameters Cenelec US D5263 Terminals IIC A B Co Ca 0 3 µF Ci Ci device C cable 13 14 15...

Page 5: ...t shall be installed in an enclosure that provides a minimum ingress protection of IP54 in accordance with IEC 60079 0 The enclosure must have a door or cover accessible only by the use of a tool The end user is responsible to ensure that the operating temperature of the module is not exceeded in the end use application Units must be protected against dirt dust extreme mechanical e g vibration imp...