1. Osicus Audiometer Important Safety Information
Read this Operating Manual before attempting to use the
Instrument.
Warnings
This Instrument is for indoor use only and it should only be used as described in this
manual.
The system must not be used in the presence of flammable gases or in an environment,
which is susceptible to explosions. (Beware of oxygen, dust and anaesthetic gases)
This unit is powered by the USB from a PC or laptop. It is advisable not to touch the
patient while using the equipment.
The equipment should be positioned in such a way that it can be easily disconnected from
the mains supply. The operation of the system can be safely terminated by switching off
or removing the mains plug.
If your desktop PC or printer does not have a power supply approved for a patient
environment, then an isolation transformer, which is in compliance with BS EN 60601,
should be used to power the PC, printer and the Osicus. You must use the isolation
transformer to ensure that the Osicus is in compliance with BS EN 60601.
Applied Parts. The Applied Parts are the headset and the response button.
Connect only items that have been specified as part of, or specified as being compatible
with the Osicus Audiometer. Supplied and compatible spare parts and accessories are
listed in Section 15 & 16 of this manual.
“
WARNING The use of PC and Printer in the Patient Environment is defined below and
must be followed to ensure safety compliance of the instrument.
”