Asra Curve User Guide V9
27/05/2021
Page
40
of
40
Instruments for dismantling and disposal, while your PC/Printer should be returned to
your supplier for dismantling and disposal.
•
These parts are subject to the WEEE directive, and should not be disposed of in landfill.
•
The only parts which are interchangeable by service personnel are the power module
and the mains cable. If these are changed, a Medical Electrical Safety test to the
current standard should be made prior to allowing release for use.
•
Circuit diagrams, component lists and parts lists are available on request, along with
email/telephone advice to service personnel trained and qualified to work on ME
devices. Modification of the ASRA Curve Audiometer is not allowed.