Globe Scientific GCC-E User Manual Download Page 3

1

Important Safety Information 

Common Safety Precautions

Carefully read the following safety precautions for a thorough understanding.

• Follow the instructions and procedures described in this manual to operate this centrifuge safely.

• Carefully read all safety messages in this manual and the safety instructions on the centrifuge.

• Safety messages are labeled as indicated below. They are in combination with signal words of “WARNING” and 

“CAUTION” with the safety alert symbol  

  to call your attention to items or operations that could be dangerous to 

you or other persons using this centrifuge. The definitions of signal words are as follows:

WARNING:

 Personal Danger

Warning notes indicate any condition or practice, which if not strictly observed, could result in personal injury or 
possible death.

CAUTION: 

Possible Damage to Centrifuge

Caution notes indicate any condition or practice, which if not strictly observed or remedied, could result in 
damage or destruction of the centrifuge.

NOTE: Notes indicate an area or subject of special merit, emphasizing either the product’s capability or common 
errors in operation or maintenance.

• Do not operate the centrifuge in any manner not described in this User Manual. When in doubt or if you have any 

troubles with this centrifuge, please CONTACT GLOBE SCIENTIFIC AND ASK FOR HELP.  

• The precautions described in this User Manual are carefully developed in an attempt to cover all the possible risks. 

However, it is also important that you are alert for unexpected incidents. Be careful operating this centrifuge.

WARNING

•  This centrifuge is not explosion-proof. Never use explosive or flammable samples.

•  Do not install the centrifuge in or near places where inflammable gases are generated or chemicals are stored.

•  Do not place dangerous materials within 30cm (approximately 12 inches) of the centrifuge.

•  Prepare all necessary safety measures before using samples that are toxic, radioactive or contaminated with 

pathogenic micro-organisms. Use of these materials are at your own risk.

•  If the centrifuge, rotor and accessories have been contaminated by solutions with toxic, radioactive or 

pathogenic materials, clean them according to the decontamination procedure as specified.

•  If you require service, please sterilize and decontaminate the centrifuge in advance and then notify the service 

center the details of the materials and procedure. 

•  To avoid electrical shocks, ensure hands are dry before handling the power cord or turning on/off the power 

switch.

•  For safety purposes, do not come within 30cm (approximately 12 inches) around the centrifuge when in 

operation.

•  While the rotor is rotating, never release the lid lock.

•  Unauthorized repairs, disassembly, or modifying the centrifuge except by an authorized Globe Scientific service 

center or service technician is strictly prohibited.

Summary of Contents for GCC-E

Page 1: ...l Centrifuge Enhanced Feature Low Speed Clinical Centrifuge USER MANUAL GCC E www globescientific com IMPORTANT Carefully and completely read this user manual to ensure user safety and achieve most ef...

Page 2: ...ps 10 11 10 1 How to Open the Lid 11 11 Rotor Instructions 12 13 11 1 Rotor Structure 12 11 2 Available Rotors and Adapters 12 12 Calculating RCF 13 13 Returning and Disposal 13 13 1 Returning the Uni...

Page 3: ...ASK FOR HELP The precautions described in this User Manual are carefully developed in an attempt to cover all the possible risks However it is also important that you are alert for unexpected inciden...

Page 4: ...he service center of the warning code if displayed along with the model and serial numbers and a detailed description of the issue Vibrations are likely to damage the centrifuge Please contact our ser...

Page 5: ...uant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment gener...

Page 6: ...d installing on a slippery surface or a surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if the ambient temperature exceeds 30 C 3 Kee...

Page 7: ...s to start the centrifuge Press to stop if centrifuge is already running Adjustment Knob Rotate clockwise to increase parameter values Rotate counterclockwise to decrease parameter values Press to swi...

Page 8: ...sure centrifuge longevity When any tube is loaded a tube of the same type volume and weight must be added in the opposite rotor location Never intentionally run the centrifuge in an unbalanced conditi...

Page 9: ...nimum Rotate the Adjustment Knob counter clockwise to change settings from large small minimum maximum 2 Set the time Press the Select button the time value flashes in the time setting mode Rotate the...

Page 10: ...speed unit g the speed symbol will then flash into RCF value input status If no button is pressed after the speed value has flashed after 5 seconds the input mode will be shut down Rotate the Adjustme...

Page 11: ...may leak into the drive units and cause corrosion or deterioration to the bearings If the rotor chamber needs cleaning clean with cloth or sponge moistened with a neutral detergent solution Sterilize...

Page 12: ...and the operator can determine the malfunction with the alarm code below 9 Maintenance Cont 9 2 Rotor Installation 1 Installation 2 Adjustment 1 2 3 4 IMPORTANT Before locking the rotor rotate it and...

Page 13: ...operation E 06 The set parameter exceeds the Modify the set parameter Incorrect allowable range value Parameter E 10 86 Refer to service manual Contact Globe Scientific 10 1 How to Open the Lid WHEN...

Page 14: ...Dimension Adapters Maximum Radius Maximum Type Type Per Rotor L mm Required Speed rpm cm RCF g 15mL Conical 8 17 120 None 4500 11 2490 1 5 5mL 12 13 82 CC AD1 4500 9 8 2218 Vacuum CC PLUG 4 7mL 13 106...

Page 15: ...be disposed of as part of domestic waste The device belongs to group 8 medical devices and is categorized in the business to business field The icon of the crossed out trash can shows that the device...

Page 16: ...on 5 Failures caused by using non standard spare parts or accessories and or unauthorized modification of the rotor or centrifuge 6 Failures caused by natural disasters including fire earthquakes etc...