USER MANUAL
BEAT-MV/BEAT-2
Edition: January 2016
- MODE OF USE
Revision: No. 1
16
3.1.3 Unit connection:
Connect the unit to a 230 V ± 15%/50-60 Hz mains source (or to
a 110 V, according to market requirement) of alternating current.
For generator group specifications, please refer to CHAPTER 6:
TECHNICAL CHARACTERISTICS.
IMPORTANT NOTES !
Connection to a generator: the generator electric connection where
the control box mains is plugged must be normalized and fitted with
differential and ground pin. Refer to the generator’s User Manual.
Connection to the mains: the building electric installation where the
control box mains is plugged must be fitted with earth connection as
well as circuit breaker.
Do not unplug the mains pulling from the cable.
BEAT-2: connect the scanner before switching on the unit.
Checking the fitting connection cable terminals before starting up the
unit is advised. There are different types of adopters depending on the
type pf fitting to be used.
Ø
Set the master switch in the ON position.
BEAT-XX No.xxxx
V v.vv UK dd/mm/yy
The display backlight will be enable, an acoustic signal
heard and the following information will appear:
BEAT-XX:
model, BEAT-MV or BEAT-2
No.xxxx:
unit serial number
V v.vv:
software version
UK:
Letters of the active language
dd/mm/yy: current date
Once the display time has elapsed (3 seconds), the unit will display the following screen.
Possible initial messages:
TECHNICAL MACHINE SERVICE
If the programmed warning date for the Technical Machine Service in the unit has expired,
the display will show:
REV. VALID UNTIL
dd/mm/yy
The unit will emit an acoustic signal and a message will
appear in the display along with the date of the next check.
If the date dd/mm/yy has not expired yet, press START to confirm and proceed to the
next screen (regardless of the existing locking set up of the unit).