Glas-Col DigiTrol II Operation Manual Download Page 7

 

12/22/14 

                                                          7

                                                                           

Rev.5 (EC 02.293.435) 

5.  When tuning is complete, the lower display indicates setpoint only and Aut reverts to OFF. The control installs 

appropriate PID tuning parameters and retains them in the non-volatile memory. 

 

 

*To abort auto-tuning, the operator must reset Aut to OFF. Auto-tuning may also be aborted by cycling power off and on.  
In all cases, aborting auto-tuning restores all original factory default values. 
 

                       

 

 

Temperature overshoot 

 
In most cases this is due to a small sample size being heated with a high wattage heater. The control is set at the factory 
with the Power Scale High Output 1 (o.hi1) setting with a value of 100. This value can be adjusted in the setup menu, 
reference page 5 for instructions on how to access and navigate through the setup menu. If excessive temperature 
overshoot is occurring, adjust the value to 50 and start the heating process from ambient room temperature. If more than 
desired temperature overshoot is still occurring, decrease to a lower value and repeat heating process from ambient room 
temperature. If the proce

ss temperature doesn’t achieve the set point temperature or takes more than desired amount of 

time to reach set point temperature, increase the power scale high output 1 value. This process can take several value 
changes to obtain proper temperature control with your particular application setup. It is recommended to initiate the Auto-
Tune process again after the proper power scale high output 1 value has been determined for your application.    

 

Error Code Definitions and Actions 

 

Indication 

Description 

Possible Causes 

Corrective Action 

 

Sensor does not provide 

a valid signal to controller 

Sensor improperly wired or open, 

Incorrect setting or sensor type, or 

Calibration corrupt 

Correct wiring or replace 

sensor, Match setting to 

sensor , or Check 

calibration of controller 

 

Controller is auto-tuning 

the control loop 

User started the auto-tune function 

Wait until auto-tune 

completes or disable 

auto-tune feature 

 

Controller is ramping to 

new set point 

Ramping feature is activated 

Disable ramping feature if 

not required 

 

Sensor input above high 

alarm set point 

Temperature is greater than alarm set 

point 

Over temperature, or Set 

alarm source to proper 

setting 

 
 
GLOSSARY

 

Automatic prompts:  Data entry points where a microprocessor-based control "prompts" or asks the operator for 
information input. 

Summary of Contents for DigiTrol II

Page 1: ...ingle output It features an auto tuning function that allows automatic setting of control parameters with a minimum of user input required This family of control accepts a type J K or T thermocouple i...

Page 2: ...finity key to display the set point temperature and the process temperature temperature at the sensor To begin the heating process use the Up Down arrow keys to enter the desired set point temperature...

Page 3: ...99 99 minutes per repeat Rate or Derivative 0 00 to 9 99 minutes Operator Interface Advance Infinity Up and Down keys and ON OFF switch 2 4 character LED display Thermocouple receptacle and 3 wire loa...

Page 4: ...Heat Proportional Band Default 25 0 F or 14 0 C C Pb Cool Proportional Band Default 25 0 F or 14 0 C ti Time Integral Default 180 td Time Derivative Default 0 o tb1 Time Base Output 1 Default 1 0 or...

Page 5: ...tion of Output 2 Default oFF h Ag Heat Algorithm Default PID hSC Hysteresis Heat Cool Default 3 0 F or 2 0 C C Ag Cool Algorithm Default oFF A ty Alarm Type Default oFF A hy Alarm Hysteresis Default 1...

Page 6: ...At and heat algorithm h Ag page 5 is set for PID mode The heat algorithm setting is set for PID from the factory When the auto tune sequence has started the lower display will flash between tun1 attn...

Page 7: ...sn t achieve the set point temperature or takes more than desired amount of time to reach set point temperature increase the power scale high output 1 value This process can take several value changes...

Page 8: ...and full Off above set point Output Action in response to difference between set point and process variable Overshoot Condition where temperature exceeds steeping due to initial power up or process ch...

Page 9: ...G 15 Amp 125 VAC less than 3 meters in length 120volt controls only Fuse Use only 250Volt fuse 15 amp rating 120volt controls only Warranty See the current Glas Col warranty policy located under the G...

Reviews: