Control mode operation

 

Manual operation provides direct (time proportioned % time) control from -100% to 100%.  A negative value is allowed 
only when Ot 1=Cool.  Automatic operation provides sensory feedback ON/OFF or PID control.  When the operation 
transfers from automatic to manual operation, the power level from automatic operation is retained and restored to the 
previous set point. 
 
The % LED indicates manual operation.  The LED is on when in Manual operation and off when in AUTO operation.  
When the LED is flashing, press the 

Infinity

 key again within 5 seconds to complete the change in operation.  It the 

sensor is open and LOC=0, 1 or 2, the control switches to Manual operation if the output was stable before the break 
occurred. 
 
 

Tuning-Automatic 

 
It is recommended that auto-tuning be used each time a new process is initiated. If the process is being duplicated, the 
parameters learned in the initial auto tune setup are stored in the memory of the control and the auto-

tune procedure isn’t 

required in most applications. 
 
Auto-tuning: This control can automatically tune the PID parameters to fit the characteristics of your particular thermal 
system.  
 
Auto-tune can only be used when Output 1 setting is hEAt and heat algorithm (h.Ag, page 5) is set for PID mode. The 
heat algorithm setting is set for PID from the factory. When the auto-tune sequence has started, the lower display will 
flash between tun1-attn and the main operation screen. 
 
Changing the set point, while in auto-tune mode, will not change the set point value. The displayed value will change but 
the control has locked in the original set point value tell the auto tune in complete or has been turned off by the user. Once 
the control has finished "learning" the system, it returns to standard PID control with the values automatically set. Auto-
tuning is complete within 80 minutes.  In order for the control to successfully complete auto-tune, the process must cross 
the 90% set point value four times within 80 minutes after auto-tune has started. If this does not happen within the 80 
minute time limit, Pb1 remains at 0 and the control will operate with the factory default PID values. 
 

 

To start auto-tuning: 

 

 

1.  Set the set point value with the arrow keys to the desired temperature of operation. 
2.  Press the 

Advance

 key until the Aut prompt appears in the lower display. 

3.  Change the parameter to 

YES 

with the arrow keys. 

4.  Press the 

Infinity

 key to display the set point and sensor temperatures. While the control is in the auto-tuning 

mode, the display will alternate between the set point and the "Aut" prompt. When tuning is complete, the lower 
display indicates only the set point. 

Summary of Contents for DigiTrol II

Page 1: ...ingle output It features an auto tuning function that allows automatic setting of control parameters with a minimum of user input required This family of control accepts a type J K or T thermocouple i...

Page 2: ...finity key to display the set point temperature and the process temperature temperature at the sensor To begin the heating process use the Up Down arrow keys to enter the desired set point temperature...

Page 3: ...99 99 minutes per repeat Rate or Derivative 0 00 to 9 99 minutes Operator Interface Advance Infinity Up and Down keys and ON OFF switch 2 4 character LED display Thermocouple receptacle and 3 wire loa...

Page 4: ...Heat Proportional Band Default 25 0 F or 14 0 C C Pb Cool Proportional Band Default 25 0 F or 14 0 C ti Time Integral Default 180 td Time Derivative Default 0 o tb1 Time Base Output 1 Default 1 0 or...

Page 5: ...tion of Output 2 Default oFF h Ag Heat Algorithm Default PID hSC Hysteresis Heat Cool Default 3 0 F or 2 0 C C Ag Cool Algorithm Default oFF A ty Alarm Type Default oFF A hy Alarm Hysteresis Default 1...

Page 6: ...At and heat algorithm h Ag page 5 is set for PID mode The heat algorithm setting is set for PID from the factory When the auto tune sequence has started the lower display will flash between tun1 attn...

Page 7: ...sn t achieve the set point temperature or takes more than desired amount of time to reach set point temperature increase the power scale high output 1 value This process can take several value changes...

Page 8: ...and full Off above set point Output Action in response to difference between set point and process variable Overshoot Condition where temperature exceeds steeping due to initial power up or process ch...

Page 9: ...G 15 Amp 125 VAC less than 3 meters in length 120volt controls only Fuse Use only 250Volt fuse 15 amp rating 120volt controls only Warranty See the current Glas Col warranty policy located under the G...

Reviews: