Guidance  and  manufacturer's  declaration  -  Electromagnetic  Immunity

The  GL-01  is  intended  for  use  in  the  electromagnetic  environment  specified  below.
The  customer  or  the  user  of  the  GL-01  should  assure  that  it  is  used  in  such  an  environment.

IMMUNITY  test

IEC  60601  test  level

Compliance  level

Electromagnetic  environment

Electrostatic 
discharge  (ESD)
IEC  61000-4-2

±  6  kV  contact
±  8  kV  air

±  6  kV  contact
±  8  kV  air

Floors  should  be  wood,  concrete  or 
ceramic  tile.  If  floors  are  covered  with 
synthetic  material,  the  relative  humidity 
should  be  at  least  30  %.

Electrical  fast 
transient/burst
IEC  61000-4-4

±  2  kV  for  power
supply  lines
±  1  kV  for 
input/output  lines

±  2  kV  for  power
supply  lines
±  1  kV  for 
input/output  lines

Mains  power  quality  should  be  that  of 
a  typical  commercial  or  hospital 
environment.

Surge
IEC  61000-4-5

±  1  kV  line(s)  to
line(s)
±  2  kV  line(s)  to 
earth

±  1  kV  differential
mode
±  2  kV  common 
mode

Mains  power  quality  should  be  that  of 
a  typical  commercial  or  hospital 
environment.

Voltage  dips, 
short  interruptions 
and  voltage 
variations  on 
power  supply 
input  lines
IEC  61000-4-11

<5  % 

U

T

(>95  %  dip  in 

U

T)

for  0,5  cycle
40  % 

U

T

(60  %  dip  in 

U

T)

for  5  cycles
70  % 

U

T

(30  %  dip  in 

U

T)

for  25  cycles
<5  % 

U

T

(>95  %  dip  in 

U

T)

for  5  s

<5  % 

U

T

(>95  %  dip  in 

U

T)

for  0,5  cycle
40  % 

U

T

(60  %  dip  in 

U

T)

for  5  cycles
70  % 

U

T

(30  %  dip  in 

U

T)

for  25  cycles
<5  % 

U

T

(>95  %  dip  in 

U

T)

for  5  s

Mains  power  quality  should  be  that  of 
a  typical  commercial  or  hospital 
environment.  If  the  user  of  the  GL-01 
requires  continued  operation  during 
power  mains  interruptions,  it  is 
recommended  that  the  GL-01  be 
powered  from  an  uninterruptible  power 
supply  or  a  battery.

Power  frequency 
(50/60  Hz) 
magnetic  field
IEC  61000-4-8

3  A/m

3  A/m

Power 

frequency 

magnetic 

fields 

should  be  at  levels  characteristic  of  a
typical 

location 

in 

typical 

commercial  or  hospital  environment.

Conducted  RF

IEC  61000-4-6

Radiated  RF

IEC  61000-4-3

3  Vrms

150  kHz  to  80  MHz

3  V/m

80  MHz  to  2,5  GHz

3  Vrms

3  V/m

Portable  and  mobile  RF 
communications  equipment  should  be 
used  no  closer  to  any  part  of  the 
GL-01,  including  cables,  than  the 
recommended  separation  distance 
calculated  from  the  equation 
applicable  to  the  frequency  of  the 
transmitter.
Recommended  separation  distance

  80  MHz  to  800  MHz
  800  MHz  to  2,5  GHz

where 

is  the  maximum  output 

power  rating  of  the  transmitter  in 
watts  (W)  according  to  the  transmitter 
manufacturer  and 

is  the 

recommended  separation  distance  in 
metres  (m).
Field  strengths  from  fixed  RF 
transmitters,  as  determined  by  an 
electromagnetic  site  survey,

should 

be  less  than  the  compliance  level  in 
each  frequency  range.

b

Interference  may  occur  in  the  vicinity 
of  equipment  marked  with  the 
following  symbol:

NOTE  1 

U

T  is  the  a.c.  mains  voltage  prior  to  application  of  the  test  level.

NOTE  2  At  80  MHz  and  800  MHz,  the  higher  frequency  range  applies.
NOTE  3  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by 
absorption  and  reflection  from  structures,  objects  and  people.

Summary of Contents for LUV 01

Page 1: ...won si Gyeongsangnam do 51140 REPUBLIC OF KOREA Tel 82 55 213 4767 Fax 82 55 213 4767 Homepage www glinc co kr OBELIS S A Bd General Wahis 53 1030 Brussels Belgium Tel 32 2 732 59 54 Fax 32 2 732 60 0...

Page 2: ...CAUTION 3 Chapter 3 Device symbols Labels Chapter 4 Laser Lancing Device Chapter 5 Performances Chapter 6 Characteristics 6 1 Principle of operation 9 6 2 Electrical rate 9 6 3 Protection type and le...

Page 3: ...Contents Chapter 8 Routine Maintanance 13 Chapter 9 Trouble Shooting 13 Chapter 10 Supplies Warranty and Customer support 14 Chapter 11 Electromagnetic Compatibility EMC Declaration 15 Chapter 12 Dis...

Page 4: ...ammable gas mixture or metallic materials 2 Contraindication Its usage is restricted or prohibited on people with hemorrhagic diseases immunization diseases or acute diseases and others those who are...

Page 5: ...ded condition of use environment This lancet is home appliance 1 6 Caution 1 Keep or use condition see 5 10 this manual 2 Do not have to sterilize and disinfect this lancet 3 Please do not use the out...

Page 6: ...device is intended for use in collection human blood samples Follow the infection control precautions as directed by your institutional policy 6 Take caution when handing any blood infected waste 7 T...

Page 7: ...4 Chapter 3 Device symbols Labels MANUFACTURER SERIAL NUMBER REFER MANUAL EXAMINATION AGENCY DIRECT CURRENT WARNING SIGN DATE OF MANUFACTURE KEEP DRY BF TYPE APPLIED PARTS A SECOND TYPE OF EQUIPMENT...

Page 8: ...Input Battery Lithium Polymer 3 7 V 460 mAh 5V 1 0A Micro USB type GL Inc Toechon dong 82 326 BI CENTER 20 Changwondaehak ro Uichang gu Chang won si Gyeongsangnam do 51140 REPUBLIC OF KOREA OBELIS S...

Page 9: ...LED Indicate operating status Level Indicator Indicate Battery charging Laser power level and Battery status check Level setting up down button Laser Shoot button Push this button when shot is ready...

Page 10: ...output 34mJ 10 GL measuring instrument GL T 01 standard 41 000 47 000 Mode Output Level 1 Level 2 Level 3 Level 4 Level 5 ENERGY 84 of Level 5 88 of Level 5 92 of Level 5 96 of Level 5 34mJ 10 41 000...

Page 11: ...8 2 Humidity 15 to 93 R H 3 Atmospheric Pressure 700hPa to 1060hPa 2 Operation condition 1 Temperature 5 40 2 Humidity 15 to 93 R H...

Page 12: ...Release Version L1F01 Software program of GL 01 5 12 Degree of protection against the ingress of water Ordinary 5 13 Mode of operation Continuous operation Equipment not suitable for use in the prese...

Page 13: ...Type B Input AC100 240V 50 60Hz Output DC 5 0V 1A Internal battery is embedded lithium polymer DC 3 7V 460mAh 5 Laser level is selected by button 6 2 Electrical rate 1 Micro USB Charger MICRO USB Typ...

Page 14: ...arance of the device 2 Open the cap and rub the surface of the focusing lens to remove dust or foreign materials with cotton swab before use 3 Touch once the shot button to check the status of interna...

Page 15: ...el Level 1 Level 2 Level 3 Level 4 Level 5 Range Infant or children Female Soft finer skin Ordinary person Ordinary person with relatively thicker person Ordinary person thick finger skin Please set t...

Page 16: ...on for 2 to 3 seconds to power off The power is automatically off if the device is not used for 2 minutes 9 Open the aperture and rub the surface of the focusing lens and put the device in the case an...

Page 17: ...e read the use manual or use instruction before use 3 Light sensitive person should consult the doctor 4 Do not draw into the water If the device is dropped into the water with POWER OFF status dry wi...

Page 18: ...15 Chapter 8 Routine Maintenance 8 1 Routine check This device is not required a routine basically Please check only the battery status before use 5 LED shows full charged status...

Page 19: ...is not operated floating 2 Please contact the customer center even if the device is not restarted when reset button is pushed in this case the device is surely out of order 3 Please change the laser...

Page 20: ...26 BI Center National Changwon University 20 Changwondaehak ro Uichang gu Changwon city Gyengsangnam do Republic of Korea 10 4 Etc 1 Expected Service Life Approximately 5 years 2 Do not open and break...

Page 21: ...ndustrial scientific and medical radio frequency equipment EN 55011 Group 1 Class A Medical Electrical Equipment is subject in regard to the electromagnetic compatibility EMC and its special precautio...

Page 22: ...ypical commercial or hospital environment If the user of the GL 01 requires continued operation during power mains interruptions it is recommended that the GL 01 be powered from an uninterruptible pow...

Page 23: ...F disturbances are controlled The customer or the user of the GL 01 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipm...

Page 24: ...sposal of old Electrical Electronic Equipment Application in the European Union and other European countries with separate collection system This symbol indicates that this product shall not be treate...

Reviews: