18
eNgLIsh
fuse
(T 50mA L 250V)
that cannot be reached from the outside, so please contact a technician authori-
sed by the manufacturer for its replacement.
Fault type
Cause
Solution
1. The suction unit doesn’t
work
Cable is damaged
External power source failure
Replace the cable
Check the external power source
2. No aspiration
Jar Cap badly screwed down
Unscrewed the cap, then rescrew it
correctly
3. No aspiration
Lid seal not in its seat
Unscrew the cap and insert the seal
properly in its seat
4. The Vacuum power on the
patient side is either very low
or absent
a) Vacuum regulator set to minimum
b) Protection filter blocked or dama-
ged
c) Connection tubes blocked, kinked
or disconnected
d) Shut-off valve blocked or dama-
ged
e) Pump motor damaged
a) Turn the vacuum regulator clockwise
and check the value of the vacuum on the
gauge
b) Replace the filter
c) Replace or reconnect the tubes, check
the jar connections
d) Empty the jar, or disconnect the tube
from the jar and unblock the shut-off val-
ve. The unit twill only work in the upright
position
e) Refer to authorised service personnel
5. The float doesn’t close
If the cap has been washed, ensure
that the float is not partially detached Insert the float into it’s place
6. The float doesn’t close
The float it’s covered by dirty mate-
rial
Unscrewed the cap, leave the and put in
on autoclave
7. Low suction
Foam inside the jar
Fill the jar to 1/3 full of ordinary water
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 None of the remedies has achieved
the desired results
Contact the seller or GIMA After-sales
Assistance Service
If the overfill security system it’s activated, don’t proceede with the liquid aspiration.
If the overfill security system doesn’t work there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological
filter who avoid the liquid penetration inside the device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the
device, in this case return the device to GIMA technical service.
GIMA Srl will provide upon request electric diagrams, components list, descriptions, setting instructions
and any other information that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA TECHNICAL SERVICE.
GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING,
APPEARS TO BE TAMPERED.
INSTRUCTION FOR USE
•
The working position must be such as to allow one to reach the control panel and to have a good view
of the empty indicator, the jar and the antibacterial filter.
•
If the device is to be transported from one place to another, to prevent the liquid collection jar from
falling and consequently the liquid from spilling, removing the jar from the device is recommended.
WARNING:
For proper use, place the aspirator on a flat, stable surface in order to have the full volume
of use of the jar and better efficiency of the overflow device. The vacuum jar, during use, must be used
in vertical mode, to prevent the action of the backflow valve. If this protection is triggered, turn the device
off and disconnect the pipe connected to the vacuum jar (indicated with the word VACUUM) on its cover.