background image

18

eNgLIsh

fuse 

(T 50mA L 250V)

 that cannot be reached from the outside, so please contact a technician authori-

sed by the manufacturer for its replacement.

Fault type

Cause

Solution

1. The suction unit doesn’t 

work

Cable is damaged  

External power source failure

Replace the cable 

Check the external power source

2. No aspiration

Jar Cap badly screwed down

Unscrewed the cap, then rescrew it 

correctly

3. No aspiration

Lid seal not in its seat

Unscrew the cap and insert the seal 

properly in its seat

4. The Vacuum power on the 

patient side is either very low 

or absent

a) Vacuum regulator set to minimum
b) Protection filter blocked or dama-

ged
c) Connection tubes blocked, kinked 

or disconnected
d) Shut-off valve blocked or dama-

ged
e) Pump motor damaged

a) Turn the vacuum regulator clockwise

and check the value of the vacuum on the

gauge

b) Replace the filter

c) Replace or reconnect the tubes, check

the jar connections

d) Empty the jar, or disconnect the tube

from the jar and unblock the shut-off val-

ve. The unit twill only work in the upright

position

e) Refer to authorised service personnel

5. The float doesn’t close

If the cap has been washed, ensure 

that the float is not partially detached Insert the float into it’s place

6. The float doesn’t close

The float it’s covered by dirty mate-

rial

Unscrewed the cap, leave the and put in 

on autoclave

7. Low suction

Foam inside the jar

Fill the jar to 1/3 full of ordinary water

Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 None of the remedies has achieved 

the desired results

Contact the seller or GIMA After-sales 

 

Assistance Service

If the overfill security system it’s activated, don’t proceede with the liquid aspiration.

If the overfill security system doesn’t work there are two cases:

1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological 

filter who avoid the liquid penetration inside the device.

2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the 

device, in this case return the device to GIMA technical service.
GIMA Srl will provide upon request electric diagrams, components list, descriptions, setting instructions 

and any other information that can help the technical assistance staff for product repair.

 BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, 

PLEASE CONTACT GIMA TECHNICAL SERVICE.

GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, 

APPEARS TO BE TAMPERED. 

INSTRUCTION FOR USE

The working position must be such as to allow one to reach the control panel and to have a good view

of the empty indicator, the jar and the antibacterial filter.

If the device is to be transported from one place to another, to prevent the liquid collection jar from

falling and consequently the liquid from spilling, removing the jar from the device is recommended.

WARNING:

 For proper use, place the aspirator on a flat, stable surface in order to have the full volume 

of use of the jar and better efficiency of the overflow device. The vacuum jar, during use, must be used 

in vertical mode, to prevent the action of the backflow valve. If this protection is triggered, turn the device 

off and disconnect the pipe connected to the vacuum jar (indicated with the word VACUUM) on its cover.

Summary of Contents for TOBI HOSPITAL

Page 1: ...ASPIRATEUR TOBI HOSPITAL CHIRURGISCHER ABSAUGER TOBI HOSPITAL ASPIRADOR QUIR RGICO TOBI HOSPITAL 28202 28232 28233 28234 M28202 M Rev 7 01 21 Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gima...

Page 2: ...unstable surfaces which may cause it to acci dentally fall and lead to a malfunction and or breakage Should there be signs of damage to the plastic parts which may expose inner parts of the energised...

Page 3: ...ive and its normatives Under certain failure conditions the temperature of the casing may become hot and there may be a risk of burns if you touch those parts In any case the temperatures do not excee...

Page 4: ...anteed up to 30 cycles of cleaning and sterilization Beyond this limit the physical chemical characteristics of the plastic material may show signs of decay Therefore we recommend that you to change i...

Page 5: ...nd cooling at environment temperature of the parts make sure that these are not damaged The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121 C 1 bar relative pressure...

Page 6: ...After sales Assistance Service If the overfill security system it s activated don t proceede with the liquid aspiration If the overfill security system doesn t work there are two cases 1 case If the...

Page 7: ...e jar is reached there is the activation of the security float the float close the aspiration connector on the jar to avoid liquid penetration inside the device The device must be used on a plan of ho...

Page 8: ...cover bends towards the inside of the glass the system is not leaking Start the aspiration and periodically check the filling level of the container The overflow valve will cause the interruption of a...

Page 9: ...it s used in such an environment Immunity Test Level indicated by the EN 60601 1 2 Compliance Level Electromagnetic environments guidance Electrostatic discharge ESD EN 61000 4 2 8kV on contact 15kV i...

Page 10: ...fixed RF transmitters as determined by an electromagnetic site study of the site should be less that the com pliance level in each frequency range Interference may occur in the vicinity of equipment...

Page 11: ...3 10 3 8 3 8 7 3 100 12 12 23 For transmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to t...

Page 12: ...specifications may change without notice Disposal The product must not be disposed of along with other domestic waste The users must dis pose of this equipment by bringing it to a specific recycling p...

Reviews: