24

 

electromagnetic interference, such as radio transmitters, mobile telephones, 
etc. Keep them far away.

 

CAUTION

: The user must check that the equipment does not have visible 

evidence of damage that may affect patient safety or monitoring capability 

 

before use.

 

The recommended inspection interval is once per month or less. If damage 

 

is evident, replacement is recommended before use.

 

CAUTION

: The following safety checks should be performed once every 

two years or as specified in the institution’s test and inspection protocol 

 

by a qualified person who has adequate training, knowledge, and practical 
experience to perform these tests. 

• Inspect the equipment for mechanical and functional damage.
• Inspect the safety relevant labels for legibility.
• Verify that the device functions properly as described in the instructions for 
use. 
• Test the patient leakage current according to IEC 60601-1: Limit: 10 uA (CF). 
The leakage current should never exceed the limit. The data should be recorded 
in an equipment log. If the device is not functioning properly or fails any of the 
above tests, the device has to be repaired.
 

CAUTION

: The battery must be properly disposed: The battery must be 

properly disposed according to local regulation

 

CAUTION

: The device shall only be used if the battery cover is closed. 

Battery must be stored in cool and dry piace.

 

If use rechargeable battery, to insure capability and life, please fully charge 
batteries before first use, normally, batteries must be continuously charged 
over 14 hours or charged according to the guidance displayed on the bat-
tery.

 

CAUTION

: Please don’t set anode and cathode of the battery wrongly. 

 

CAUTION

: The valid period of this product is five years.

 

After the service life, please return the products to the manufacture or 
disposeal the products according to local regulations.

 

When cleaning the machine:

 

CAUTION

: Don’t use strong solvent, for example, acetone.

 

CAUTION

: Never use an abrasive such as steel wool or metal polish.

 

CAUTION

: Do not allow any liquid to enter the product, and do not 

 

immerse any parts of the device into any liquids.

 

CAUTION

: Avoid pouring liquids on the device while cleaning.

 

CAUTION

: Don’t remain any cleaning solution on the surface of the device. 

ENGLISH

Summary of Contents for Sonoline C 29480

Page 1: ...o ATTENTION The operators must carefully read and completely understand the present manual before using the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le pr...

Page 2: ...operations repairs are carried out by persons authorized by the manufacturer and the device is used in accordance with the instructions for use WARNING This device is not intended for treatment The i...

Page 3: ...equipment Furthermore all configurations shall comply with the valid version of the system standard IEC60601 1 1 Everybody who connects additional equipment to the signal input connector or signal out...

Page 4: ...evice is not functioning properly or fails any of the above tests the device has to be repaired CAUTION The battery must be properly disposed The battery must be properly disposed according to local r...

Page 5: ...alculating unit LCD display control unit etc It has 3 work modes real time FHR display mode averaged FHR display mode and manual mode It also has audio output and can be connected with earphone or rec...

Page 6: ...26 Chapter 3 Outlook and Configuration Fig 3 1 Front Panel Fig 3 2 Rear Panel Start Stop Button Mode Button Power Button Probe LCD ENGLISH Speaker Battery Compartment...

Page 7: ...volume control button on Pocket Fetal Doppler The primary functions are as follows 3 2 1 Power Button Function Power on off Power on Push the button once Power off Push down the button and hold 3 sec...

Page 8: ...dio output and can be connected with earphone or recorder with audio input to record The socket terminal post or switch that connected with the headphones Accessory equipment connected to the analog a...

Page 9: ...al heart Apply a liberal amount of gel to the faceplate of probe place the faceplate of probe at the best position for detecting fetal heart Adjust the probe to obtain an optimum audio signal ideally...

Page 10: ...l heart rate bar graph on LCD indicates the strength of the fetal heart rate signals the shown fetal heart rate values and heartbeat waveform changes slowly 4 2 3 Manual Mode Mode 3 Press the start st...

Page 11: ...avoid falling off stress etc When the device is not used for a long time users are recommended to connect the plug of one probe to device socket and put the probe in the parking Then pack the device w...

Page 12: ...checks to ensure the fetal safety 4 5 Battery Status Indicator When it works normally the LCD screen displays the status of the battery as follows Battery power is full Battery power is not full Batt...

Page 13: ...Fig 4 2 Fig 4 2 Replacing Battery 2 Put two AA size batteries into the battery compartment as for the direction of battery please refer to the instruction inside the battery compartment at last close...

Page 14: ...wered equipment Anti electroshock Degree Type CF applied part Harmful Liquid Proof Degree Main unit Degrees of protection provided by enclosure IPXO Probe Prevent from water splashing degree of protec...

Page 15: ...signal power off automatically Battery Type Recommended Two pieces of 1 5 V DC battery SIZE AA LR6 Probe Nominal Frequency 2 0MHz Working Frequency 2 0 MHz 10 P 1 MPa Iob 20 mW cm2 Ispta 100 mW cm2 Ul...

Page 16: ...ree of dust and dirt clean exterior surface display screen included of the chassis with a dry soft cloth If necessary clean the chassis with a soft cloth soaked in a solution of soap or water and wipe...

Page 17: ...ice please solve them as below Problems Possible reasons Solutions ENGLISH Weak sound Noise Low sensitivity Volume is too low Power is low Did not daub the gel Probe is too near from the main unt Dist...

Page 18: ...more oxygen relatively the oxygen for fetus become less It is easy to arose symptoms such as fetus anoxia Testing the FHR at this time can display the healthy status for the fetus best The above thre...

Page 19: ...4 8 3 4 B a 40 9 6 6 8 T 20 4 8 1 B a 40 9 6 2 B w dB B dB V s r m s mV V n r m s mV 0 57 6 186 94 5 93 109 2 0 56 4 175 90 5 78 107 8 0 56 4 174 89 5 82 107 9 0 49 6 173 90 5 68 100 9 0 60 8 178 89...

Page 20: ...ity of the warranty GIMA will repair and or replace free of charge all the defected parts due to production reasons Labor costs and personnel traveling expenses and packaging not included All componen...

Page 21: ...122...

Page 22: ...123...

Reviews: