6.

 

Adjustments 

6.1

 

Yearly inspections by the operator 

 

Keep to the yearly inspection schedules and inspect the product according to IEC 62353 standard. 

 

6.2

 

Repairs 

 

The Product must only be opened and repaired by a technician who has attended a course on the Product organised 
by the manufacturer or by a qualified technician in possession of the necessary technical skills. 
 

6.3

 

Adjustments 

 

The  Product  is  sold  already  balanced  and  does  not  require  further  adjustment.  In  the  event,  In  the  event  of  the 
Product becoming stiff or loose over time, please contact the service center. 
 

6.4

 

Troubleshooting 

 

Problem 

 

Solution 

The Product does not switch on 

Check the INPUT  and OUTPUT (24V) power supply voltage and 
the voltage of the electronic board terminal (24V).  

The Product does not remain in position

 

See the attached instructions

 

The light flickers

 

Contact the service center 

The light beam on the operating field is 
not focalized

 

Contact the service center 

 
 

6.5

 

Routine maintenance 

 

Period

 

Job 

Once a year

 

Inspect  all  the  lamp  joints  and  make  sure  they  are  not  hard to move. If Product does not 
maintain the position or is hard in movements, see the attached. 

Once a year

 

Make sure the retention screws are tightened properly.

 

If these are not properly fastened, 

adequately tighten.

 

Once a year

 

Check the condition of the Product paint. Make sure there are no paint pieces that could 
fall on the operating field.

 

  

 

6.6

 

Spare parts list 

 

Description 

Order code 

Sterilizable handle

 

Z100848 

Electronic board

 

Z300640 

Electronic card

 

Z300231 

 

 

 

Summary of Contents for PRIMALED

Page 1: ...ation mobile version 6 2 2 Installation wall version S 12MED fixing 6 2 3 Installation wall version rail bar fixing 6 2 4 Installation table version S11 fixing 6 2 5 First starting 6 2 6 Installation...

Page 2: ...alian edition of this operator s manual 1 General information The ME Medical Electrical EQUIPMENT to which this manual refers is a LAMP for diagnosis or observation For easier description such ME EQUI...

Page 3: ...tion manual and user manual Transport is carried out by any road haulage contractor as long as they respect the following characteristics Temperature C 15 60 Humidity 10 75 Atmospheric pressure h Pa 5...

Page 4: ...ential performance Collateral standard Electromagnetic compatibility Requierements and tests Classification with reference to article 9 and Annex IX of Directives 93 42 EEC and 2007 47 EC DURATION Sho...

Page 5: ...ot above 40 of the net value of the product as indicated on the buyer s invoice GIMA s liability is expressly ruled out for indirect damages or consequential damages including cases of the lamp not be...

Page 6: ...the bar and tighten the handle 3 Insert the lamp into the upper hole of the rail clamp 4 Follow points 2 and 3 of paragraph 2 1 2 4 Installation table version S11 fixing 1 Fix the clamp S 11 to the ta...

Page 7: ...ions could cause glare and potential damage to the retina Never place and or hang anything on the Product Unless this precaution is taken positioning will not be reliable and the danger exists of such...

Page 8: ...ic parts becoming fragile use only disinfectants with low alcohol content Dose the disinfectants so no liquids penetrate inside the lamp bodies and into the support arm system Clean the Product with a...

Page 9: ...tch on Check the INPUT and OUTPUT 24V power supply voltage and the voltage of the electronic board terminal 24V 2 The Product does not remain in position See the attached instructions 3 The light flic...

Page 10: ...240 Frequency Hz 50 60 Absorbed power VA 30 Light source N 12 Led x 1 4W LED light source duration h this datum can vary according to voltage peaks and the frequency of use 50 000 Light intensity con...

Page 11: ...Operative Manual MO079 IT 15 06 12 Rev 0 Pag 11 di 16 8 Electrical diagram...

Page 12: ...ngs used for domestic purposes provided the following warning is heeded Warning This equipment system is intended for use by healthcare professionals only This equipment system may cause radio interfe...

Page 13: ...cial or hospital environment NOTE UT is the a c main voltage prior to application of the test level Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted R...

Page 14: ...transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 24 0 1 0 38 0 38 0 73 1 1 2 1...

Page 15: ...to adjust the friction force in correspondence of rotational joints and therefore the consequent movement of movable arms ROTATION JOINTS Depending on the version of the device there is a different n...

Page 16: ...Operative Manual MO079 IT 15 06 12 Rev 0 Pag 16 di 16 Note...

Reviews: