11 

6.2.    Attention for operation 

A.

    Please check the device before using, and confirm that it can work normally. 

B

. The finger should be in a proper position (see the attached illustration of figure 4 for reference), or 

else it may result in inaccurate measure. 

C. 

The SpO

2

 sensor and photoelectric receiving tube should be arranged in a way with the subject’s 

arteriole in a position there between. 

D.

 The SpO

2

 sensor should not be used at a location or limb tied with arterial canal or blood pressure 

cuff or receiving intravenous injection. 

E.

 Do not fix the SpO

2

 sensor with adhesive or else it may result in venous pulsation and inaccurate 

measure of SpO

and pulse rate.

 

F.

 Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby 

light, infrared heater, direct sunlight and etc. 

G.

  Strenuous  action  of  the  subject  or  extreme  electrosurgical  interference  may  also  affect  the 

accuracy. 

H.

 Testee can not use enamel or other makeup. 

I.

 Please clean and disinfect the device after operating according to the User Manual (7.1).

 

6.3.    Clinical restrictions 

A.

 As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject 

is  required.  For  a  subject  with  weak  pulse  due  to  shock,  low  ambient/body  temperature,  major 

bleeding,  or  use  of  vascular  contracting  drug,  the  SpO

2

  waveform  (PLETH)  will  decrease.  In  this 

case, the measurement will be more sensitive to interference. 

B.

  For  those  with  a  substantial  amount  of  staining  dilution  drug  (such  as  methylene  blue,  indigo 

green and acid indigo blue), or carbon monoxide hemoglobin (COHb),  or methionine (Me+Hb) or 

thiosalicylic  hemoglobin,  and  some  with  icterus  problem,  the  SpO

2

  determination  by  this  monitor 

may be inaccurate. 

C.

 The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor 

blamed for serious error of SpO

2

 measure. 

D. 

As the SpO

2

 value serves as a reference value for judgement of anemic anoxia and toxic anoxia, 

some patients with serious anemia may also report good SpO

2

 measurement. 

7.    Maintain

transportation and storage 

7.1.    Cleaning and Disinfecting 

Using medical alcohol to disinfect the device, nature dry or clean it with clean soft cloth. 

7.2.    Maintain   

A.

 Please clean and disinfect the device before using according to the User Manual(7.1). 

B.

 Please change the battery when the screen shows 

C. 

Take out the battery if leave the equipment unused for long time. 

D. 

The device needs to be calibrated once a year (or according to the calibrating program of hospital). 

It also can be performed at the state-appointed agent or just contact us for calibration. 

Summary of Contents for OXY-50

Page 1: ...I Cod 35100 OXY 50 Pulse Oximeter User manual ...

Page 2: ...onitoring abnormality human injury and equipment damage due to users negligence of the operation instructions The manufacturer s warranty service does not cover such faults Owing to the forthcoming renovation the specific products you received may not be totally in accordance with the description of this User Manual We would sincerely regret for that This product is medical device which can be use...

Page 3: ...Installation 4 5 1 View of the front panel 4 5 2 Under side View and Left View 5 5 3 Battery and probe installation 5 5 3 Accessories 6 6 Operating Guide 6 6 1 Application method 6 6 2 Attention for operation 10 6 3 Clinical restrictions 11 7 Maintain transportation and storage 11 7 1 Cleaning and Disinfecting 11 7 2 Maintain 11 7 3 Transportation and storage 12 8 Troubleshooting 12 9 Key of Symbo...

Page 4: ...r else the device may have the possibility of working abnormally Please choose the accessories and probe which are approved or manufactured by the manufacturer or else it may damage the device The device can only be matched with the compatible probe Please don t measure this device with functional tester for the device s related information 1 3 Attention Keep the oximeter away from dust vibration ...

Page 5: ...se the decrease of SpO2 in the blood furthermore some other causes such as the malfunction of human body s self adjustment damages during surgery and the injuries caused by some medical checkup would also lead to the difficulty of oxygen supply in human body and the corresponding symptoms would appear as a consequence such as vertigo impotence vomit etc Serious symptoms might bring danger to human...

Page 6: ... two beams of different wavelength of lights can be focused onto human nail tip through perspective clamp finger type sensor Then measured signal can be obtained by a photosensitive element information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor Figure 1 4 Technical specifications 4 1 Main performance A SpO2 value display B Pulse rate ...

Page 7: ...2 and pulse rate can be shown correctly when pulse filling ratio is 0 4 SpO2 error is 4 pulse rate error is 2 bpm or 2 select larger E Resistance to surrounding light The deviation between the value measured in the condition of man made light or indoor natural light and that of darkroom is less than 1 F Power supply requirement 2 6 V DC 3 6V DC G Optical Sensor Red light wavelength is 660nm 6 65mW...

Page 8: ...Refer to Figure 4 and insert the two AA size batteries properly in the right direction Figure 4 Batteries installation B Replace the cover C Inserting the SpO2 probe of the pulse oximeter in the upper jack The probe is limited to be produced by our company never replace it with the similar ones by other manufacturers Please take care when you insert the batteries for the improper insertion may dam...

Page 9: ... e The data can be read directly from the screen on the measuring interface Fingernails and the luminescent tube should be on the same side If the alarm function is on the device will provide medium priority alarm signal when probe or finger is out Intermittent alarm will occur and the user interface presents FINGER OUT Medium priority indicating that prompt operator response is required B Pause a...

Page 10: ...ine shows the PR trend waveform the time underside shows the starting time of dispalying the date in the screen the middle and underside the screen means the operation direction of the Down button Press right button it will show in the position then press Down button to enter next hour Press left button it will show in the position then press Down button to enter last hour d The Review Interface 2...

Page 11: ...se the parameter to be adjusted then press the left button or right button to change data Each press of the left button or right button the data will raise or descend for one time accordingly If the alarm function is on the device will provide medium priority alarm signal when the data of SpO2 or pulse rate is beyond the limit Intermittent alarm will occur and the measurement shows in yellow Mediu...

Page 12: ...gure 10 if it is in recording state will come into figure 11 Figure 10 b In the status shown in Figure 10 press left button or right button can change the setting of the item then press menu button to exit the status in Figure 10 and perform setting YES for starting recording NO for do not recording c In the status shown in Figure 11 press left button or right button can change the setting press m...

Page 13: ...anges of time due to improper operation You can change the state by press the left button or right button choose yes to reset the time choose no to forbid time resetting b Press the up button or down button to select the parameter that you want to change then adjust the data by press the left button or right button c Exit the clock setting menu directly by press the menu button If you have reset t...

Page 14: ...a subject with weak pulse due to shock low ambient body temperature major bleeding or use of vascular contracting drug the SpO2 waveform PLETH will decrease In this case the measurement will be more sensitive to interference B For those with a substantial amount of staining dilution drug such as methylene blue indigo green and acid indigo blue or carbon monoxide hemoglobin COHb or methionine Me Hb...

Page 15: ... a diagnosis if you are sure the device works all right The SpO2 and Pulse Rate are not displayed stably 1 The finger is not placed inside deep enough 2 The finger is shaking or the patient is moving 1 Place the finger properly and try again 2 Let the patient keep calm The device can not be turned on 1 The battery is drained away or almost drained away 2 The battery is installed incorrectly 3 The ...

Page 16: ...on Fully charged battery Low battery Power on off button Left button Alarm pause button Menu button Right button Down button Up button USB Type BF SN Serial number 1 The finger clip falls off no finger inserted 2 Probe error 3 Signal inadequacy indicator IPX1 Ingress of liquids rank WEEE 2002 96 EC ...

Page 17: ...viation between average value and true value does not exceed 1 Pulse Parameter Specification Measuring range 30bpm 250bpm the resolution is 1bpm Accuracy 2bpm or 2 select larger Average pulse rate Moving calculate the Average pulse rate every 4 cardio beat s cycle The deviation between average value and true value does not exceed 1 Safety Type Interior Battery BFType Pulse Intensity Range Continuo...

Page 18: ...ght About 180g with Dry battery 2AA Appendix 1 State Alarm condition delay Alarm signal generation delay Low voltage alarm 1s 20ms Spo2 alarm 330ms 20ms Pulse rate alarm 330ms 20ms Probe error alarm 16ms 20ms ...

Page 19: ...ts due to production reasons Labor costs and personnel traveling expenses and packaging not included All components subject to wear are not included in the warranty The repair or replacement performed during the warranty period shall not extend the warranty The warranty is void in the following cases repairs performed by unauthorized personnel or with non original spare parts defects caused by neg...

Reviews: