background image

PROFESSIONAL MEDICAL PRODUCTS

PLEASE READ CAREFULLY BEFORE USING

This digital thermometer provides a quick and highly accurate reading of an 

individual’s body temperature. The digital thermometer is intended to meas-

ure the human body’s temperature in regular mode orally, rectally or under 

the arm, and the device is reusable for clinical or home use on people of all 

ages. To better understand its functions and to provide years of dependable 

results, please read all instructions first.

This appliance conforms to the following standards:

EN 12470-3 Clinical thermometers —Part 3: Performance of compact elec-

trical thermometers (non-predictive and predictive) with maximum device,

ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular require-

ments for basic safety and essential performance of clinical thermometers 

for  body  temperature  measurement,  EN  60601-1-11  Medical  electrical 

equipment —Part 1-11: General requirements for basic safety and essen-

tial performance – Collateral Standard: Requirements for medical electrical 

equipment and medical electrical systems used in the home healthcare en-

vironment and complies with the requirements of EN 60601-1-2(EMC), IEC/

EN60601-1(Safety) standards. And the manufacturer is ISO 13485 certified.

 Warning:

•  Read instructions thoroughly before using digital thermometer.

•  Choking Hazard: Thermometer cap and battery may be fatal if swallowed. 

Do not allow children to use this device without parental supervision.

•  Do not use thermometer in ear. Designed use is for oral, rectal, and armpit 

(axilla) readings only.

•  Do not place thermometer battery near extreme heat as it may explode.

•  Remove battery from the device when not in operation for a long time.

Note: Use of the probe cover may result in a 0.1℃(0.2℉) discrepancy 

from actual temperature.

•  The use of temperature readings for self-diagnosis is dangerous. Consult 

your doctor for the interpretation of results. Self-diagnosis may lead to the 

worsening of existing disease conditions.

•  Do not attempt measurements when the thermometer is wet as inaccurate 

readings may result.

•  Do not bite the thermometer. Doing so may lead to breakage and/or injury.

•  Do not attempt to disassemble or repair the thermometer. Doing so may 

result in inaccurate readings.

•  After each use, disinfect the thermometer especially in case the device is 

used by more than one person.

•  Do not force the thermometer into the rectum. Stop insertion and abort the 

measurement when pain is present. Failure to do so may lead to injury.

•  Do not use thermometer orally after being used rectally.

•  For children who are two years old or younger, please do not use the de-

vices orally.

•  If the unit has been stored at temperatures over 5℃~40℃(41℉~104℉), 

leave it in 5℃~40℃ (41℉~104℉) ambient temperature for about 15 min-

utes before using it.

PRECAUTION

The performance of the device may be degraded should one or more of the 

following occur:

- Operation  outside  the  manufacturer’s  stated  temperature  and  humidity 

range.

- Storage  outside  the  manufacturer’s  stated  temperature  and  humidity 

range.

- Mechanical shock (for example, drop test) or degraded sensor.

- Patient temperature is below ambient temperature.

* Portable and mobile RF communications can affect the device. The device 

needs special pre-cautions regarding EMC according to the EMC informa-

tion provided in the accompany documents.

SPECIFICATIONS

DIRECTIONS

1. Press  the  On/Off  Button  next  to  LCD  display. A  tone  will  sound  as  the 

screen shows 

, followed by last recorded temperature. After show-

ing the self-test temperature, the thermometer is now in the testing mode.

2. Position thermometer in desired location (mouth, rectum, or armpit.) 

a) Oral Use:  Place  thermometer  under  tongue  as 

indicated  by  “√”  position  shown  in  Figure  2.  Close 

your mouth and breathe evenly through the nose to 

prevent the measurement from being influenced by 

inhaled/exhaled  air.  Normal  temperature  between 

35.7℃ and 37.3℃ (96.3℉ and 99.1℉). Figure 2

b) Rectal Use: Lubricate silver probe tip with petrole-

um jelly for easy insertion. Gently insert sensor approximately 1cm (less 

than 1/2″) into rectum. Normal temperature between 36.2 and 37.7℃ 

(97.2℉ and 99.9℉).

c) Armpit Use:  Wipe  armpit  dry.  Place  probe  in  armpit  and  keep  arm 

pressed firmly at side. From a medical viewpoint, this method will al-

ways  provide  inaccurate  readings,  and  should  not  be  used  if  precise 

measurements are required. Normal temperature between 35.2℃ and 

36.7℃(95.4℉ and 98.1℉).

3. The degree sign flashes throughout  the testing process. When flashing 

stops  an  alarm  will  beep  for  approximately  10  seconds. The  measured 

reading will appear on the LCD simultaneously. The minimum measure-

ment time until the signaling tone (beep) must be maintained without ex-

ception. The measurement continues even after the buzzer notification. 

So that in order to achieve better body temperature measurement result, 

recommend to keep the probe in mouth and rectum about 2 minutes, or 

in armpit about 5 minutes regardless of the beep sound and at least 30 

seconds measurement interval should be maintained.

*Note: Normally the buzzes are“ Bi-Bi-Bi- Bi-”;Alarm beeps more rap-

idly when temperature        reaches 37.8 (100℉) or higher ,and the 

buzzes are“ Bi-Bi-Bi----- Bi-Bi-Bi------- Bi-Bi-Bi”

4. To prolong battery life, press the On/Off Button to turn unit off after testing 

is complete. If no action is taken, the unit will automatically shut off after 

around10 minutes.

TROUBLESHOOTING

Probe

ON/Off Button

LCD

Battery Cap

Error 

message

Problem

Solution

Temperature taken is

lower than 32.0°C 

(90.0°F)

Turn off, wait one minute and 

take a new temperature via 

close contact and sufficient rest.

Temperature taken is

higher than 42.9°C 

(109.9°F)

Turn off, wait one minute and 

take a new temperature via 

close contact and sufficient rest.

ENGLISH

Type:

Digital Thermometer (Not Predictive)

Measure Range:

32.0℃ - 42.9℃° (90.0℉-109.9℉ )

(℃/℉ chosen by manufacturer)

Accuracy:

±0.1℃ (±0.2℉) during 35.5℃~42.0℃ 

(95.9℉~107.6℉) at 18℃~28℃ 

(64.4℉~82.4℉) ambient operating range 

±0.2℃ (±0.4℉) for other measuring and 

ambient operating range

Operating mode:

Direct Mode

Display:

Liquid crystal display, 3 1/2 digits

Memory:

For storing the last measured value

Battery:

One 1.5 V DC. button battery 

(size LR41or SR41, UCC 392)

Battery life:

Approx. 200hours of continuous operation 

or 1 year with 3 measurements per day

Dimension:

13.9cm×2.3cm×1.3cm (L×W×H)

Weight:

Approx. 13 grams including battery

Expected service life:

Three years

Ambient operating range: Temperature: 5℃~40℃(41℉~104℉) 

Relative humidity: 15%~95%RH 

Atmospheric Pressure : 700hPa ~ 1060hPa

Storage and transporta-

tion condition:

Temperature: -20℃~55℃(-4℉~131℉) 

Relative humidity: 15%~95%RH 

Atmospheric Pressure : 700hPa ~ 1060hPa

Ingress Protection Rating: IP 27
Classification:

Type BF 