background image

BATTERY REPLACEMENT

1. Replace battery when   appears in the lower right corner of LCD display.

2. Pull battery cover off as shown in Figure 3.

3. Gently pull out plastic circuit board with battery chamber approximately 1 

cm (slightly less than1/2

4. Use pointed object such as a pen to remove old battery. Discard battery 

lawfully. Replace with new 1.5V DC button type LR41 or SR41,UCC392, 

or equivalent. Be sure battery is installed with “+” polarity facing up. (See 

Figure 5)

5. Slide battery chamber back into place and attach cover.

CALIBRATION

The thermometer is initially calibrated at the time of manufacture. If the ther-

mometer is used according to the use instruction, periodic readjustment is 

not required. However, we recommend checking calibration every two years 

or whenever clinical accuracy of the thermometer is in question. Turn on the 

thermometer and insert into the water bath and then check the laboratory 

accuracy of thermometer. Please send the complete device to the dealers 

or manufacturer.

The above recommendations do not supersede the legal requirements. The 

user must always comply with legal requirements for the control of the meas-

urement, functionality, and accuracy of the device which are required by the 

scope of relevant laws, directives or ordinances where the device is used.

CLEANING AND DISINFECTION

Wipe the thermometer with a soft clean cloth.

For stubborn stains, wipe the thermometer with a cloth that has been damp-

ened with water or a neutral detergent solution and then wring thoroughly. 

Finish by wiping with a soft dry cloth.

For disinfection, 75% Ethanol or Isopropyl alcohol can be used.

Observe the following to prevent damage to the thermometer.

Do not use benzene, thinner, gasoline or other strong solvents to clean the 

thermometer.

Do not attempt to disinfect the sensing section (tip) of the thermometer by 

immersing in alcohol or in hot water (water over 50℃ (122℉).

Do not use ultrasonic washing to clean the thermometer.

Disposal: The product must not be disposed of along with other do-

mestic waste. The users must dispose of this equipment by bringing 

it to a specific recycling point for electric and electronic equipment. 

GIMA WARRANTY TERMS

The Gima 12-month standard B2B warranty applies.

The system is not

functioning properly.

Unload the battery, wait for 

1 minute and repower it. If the 

message reappears, contact the 

retailer for service.

Dead battery: Battery 

icon is flashing, can’t be 

measurable.

Replace the battery.

Type BF applied part

Follow instructions for use

Manufacturer

Date of manufacture

Keep in a cool, 

dry place

Product code

Keep away from 

sunlight

Lot number 

Authorized represent-

ative in the European 

community

Caution: read instructions 

(warnings) carefully

WEEE disposal

Medical Device complies 

with Directive 93/42/EEC

Temperature limit 

IP27

Covering Protection rate

Humidity limit 

Direct current

Atmospheric pressure 

limit 

Reviews: