• Inspect and test each device before each use.

• Discard devices that have reached their life expectancy.

• Activation of an electrosurgical device when not in contact with target tissue or in position to deliver

energy to target tissue (fulguration) may cause capacitive coupling.

INSPECTION

These devices should be inspected before and after each use. Visually examine the devices for obvious 

physical damage including:

• Cracked, broken or otherwise distorted plastic parts.

• Broken or significantly bent connector contacts.

• Damage including cuts, punctures, nicks, abrasion, unusual lumps, significant discoloration.

• Tips for damage, corrosion or misalignment condition.

CLEANING

Remove any obvious debris accumulated during use from the device with a soft, non-metallic instrument 

cleaning brush plus mild detergent and sterile purified water solution.

Rinse thoroughly with sterile, purified water until free of detergent residual and debris, then thoroughly 

dry using a sterile wipe, (Do not fully immerse in fluids).

PRECAUTIONS

• Do not scrape or scrub coated surface with abrasives.

• Do not use cervical brush to clean smoke tube.

• Do not overlap instruments in sterilization tray.

• Do not soak coated instruments in CIDEX or other caustic cold sterilization solutions.

• Do not use bleach.

• Do not place coated instruments in ultrasonic cleaner.

STERILIZATION

Wrap each pair of instrument separately or place in a container so as to prevent instrument from contact-

ing each other or other instruments.

• Steam: Steam autoclave at 250°F (121ºC) for 20 minutes.

• Flash: Steam autoclave at 275°F (134ºC) for 10 minutes.

• Gas (ETO): Follow manufacturer’s guide for the unit which you are using.

• Pre-Vac: Steam autoclave at 270°-273°F (132°-135°C),

Exposure Time (4 minutes), Dry Time (20 minutes)

Do not handle the device until they are thoroughly cooled. For maximum life of the product ETO sterili-

zation is recommended.

STORAGE & HANDLING

Electrosurgical Cables must be stored in a clean, cool and dry area. Protect from mechanical damage & 

direct sun light. Handle with extreme care.

Index of symbols

Caution: read instructions (warnings) carefully

Follow instructions for use

Keep in a cool, dry place

Keep away from sunlight

Manufacturer

Date of manufacture

Lot number 

Product code

Medical Device complies with Directive 93/42/

EEC 

Authorized representative in the European 

community

Non-sterile

WEEE disposal

ENGLISH 

Summary of Contents for 30617

Page 1: ... ATTENZIONE Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual before using the product AVIS Les opérateurs doivent lire et bien comprendre ce manuel avant d utiliser le produit ACHTUNG Die Bediener müssen vorher dieses Handbuch gelesen und verstanden haben bevor...

Page 2: ...devices Connect the connectors to the chosen surgical device and then the feeding device and verify that the connections are correct and that no metallic parts are left exposed Do not bending the cable or fix it in any way with clips or similar devices that could damage its function ality REUSE We guarantee our products to withstand a minimum of 20 sterilization cycles when sterilized in accord an...

Page 3: ...ervical brush to clean smoke tube Do not overlap instruments in sterilization tray Do not soak coated instruments in CIDEX or other caustic cold sterilization solutions Do not use bleach Do not place coated instruments in ultrasonic cleaner STERILIZATION Wrap each pair of instrument separately or place in a container so as to prevent instrument from contact ing each other or other instruments Stea...

Page 4: ...the warranty GIMA will repair and or replace free of charge all the defected parts due to production reasons Labor costs and personnel traveling expenses and packaging not included All components subject to wear are not included in the warranty The repair or replacement performed during the warranty period shall not extend the warranty The warranty is void in the following cases repairs performed ...

Reviews: