leave the thermometer unattended in the hands of children.

• Avoid bending the thermometer tip.

• If the ambient temperature is over 35°C or 95°F, dip the thermome-

ter tip in cold water for approx. 5 to 10 seconds prior to measuring 

the temperature. Persistent fever, in particular in children, has to be 

treated by a doctor - please get in touch with your doctor!

• Do not use near strong electromagnetic fields, i.e.

Keep it away from any radio systems and mobile phones.

Battery replacement

The battery is empty and needs replacing when the 

 or 

battery symbol appears on the right of the LCD. Remove the battery 

cover and replace it with a battery (preferably non-mercury) of the 

same type. 

We advise you to remove the batteries if the device is not going to be 

used for a longer period of time.

Technical data

Type: maximum thermometer

Measurement range: (32.0~ 42.9)°C/(89.6~109.2)°F 

Measurement accuracy:

Measurement accuracy:

+/- 0.1°C/0.2°F  (35.5°C~42.0°C/95.9°F~107.6°F)

+/-0.2°C/0.4°F (32.0°C~35.5°C/89.6°F~95.9°F)

(42.0°C~42.9°C/107.6°F~109.2°F)

Storage/transportation temperature: (-25~55)°C, ≤95%RH

Ambient temperature during use: (5~35)°C, ≤80%RH

Atmosferic pressure 700~1060 hPa 

Min Scale: 0.1°C/0.1°F

Battery type: alkaline battery, type LR41, 1.5V, service life minimum 

100 hours under continuous operation. 

Explanation of symbols

Legal requirements and guidelines

This product complies with the European Directive for Medical Devic-

es 93/42/EEC and carries the CE mark. The device also complies with 

the specifications of the European Standard EN 12470-3 Clinical ther-

mometers-Part 3: Performance of compact electrical thermometers 

(non-predictive and predictive) with maximum device.

Calibration check

This thermometer is initially calibrated at the time of manufacture. If 

this thermometer is used according to the operation instruction, peri-

odic re-adjustment is not required.

The calibration check has to be carried out immediately, if there are 

indications that the product does not keep the defined error limits or 

the calibration properties could have been affected by an intervention 

or by any other means. Please also observe any national statutory 

regulations. The calibration check can be carried out by the compe-

tent authorities or by authorised service providers. A test instruction 

for calibration check can be provided to the relevant authorities and 

authorised services providers on request.

GIMA WARRANTY CONDITIONS

Congratulations for purchasing a GIMA product. This product meets 

high qualitative standards both as regards the material and the pro-

duction. The warranty is valid for 12 months from the date of supply 

of GIMA. During the period of validity of the warranty, GIMA will repair 

and/or replace free of charge all the defected parts due to production 

reasons. Labor costs and personnel traveling expenses and packag-

ing not included. All components subject to wear are not included in 

the warranty. The repair or replacement performed during the warranty 

period shall not extend the warranty. The warranty is void in the fol-

lowing cases: repairs performed by unauthorized personnel or with 

non-original spare parts, defects caused by negligence or incorrect 

use. GIMA cannot be held responsible for malfunctioning on electron-

ic devices or software due to outside agents such as: voltage chang-

es, electro-magnetic fields, radio interferences, etc. The warranty is 

void if the above regulations are not observed and if the serial code (if 

available) has been removed, cancelled or changed.

The defected products must be returned only to the dealer the prod-

uct was purchased from. Products sent to GIMA will be rejected.

Disposal:

 The product must not be disposed of along with oth-

er domestic waste. The users must dispose of this equipment 

by bringing it to a specific recycling point for electric and elec-

tronic equipment. For further information on recycling points 

contact the local authorities, the local recycling center or the 

shop where the product was purchased. If the equipment is 

not disposed of correctly, fines or penalties may be applied in 

accordance with the national legislation and regulations.

 or 

Battery is empty

WEEE disposal

Lo°C o Lo°F

Temperature under 32°C or 89.6°F

Hi°C o Hi°F

Temperature over 42.9°C or 109.2°F

Type BF applied part

Caution: read instructions (warnings) carefully

Keep away from sunlight

Keep in a cool, dry place

Follow instructions for use

Manufacturer

Medical Device complies with Directive 93/42/EEC 

Product code

Lot number

IP22

Covering Protection rate 