background image

Gilson Company, Inc. 

Vibro-Deairator: SGA-5R & SGA-5RT

Page 5

3.3 SG-15 Plastic Pycnometer:

Place the adapter ring of the SGA-7 Adapter Set between 
the four rubber bumpers on the agitator base, then posi-
tion the SGA-15 on the ring. The upper adapter of the 
SGA-7 is placed with the foam side down on top of the 
pycnometer. Slide the top plate over the adjustment rods 
and position the adapter between the four adapters (see 
Figure 3). Adjust the plastic threaded knobs until there is 
a small gap between the plate and the knobs. Drop the 
metal speed-knobs onto the rods and tighten so that the 
plate is pushed down slightly and contacts the threaded 
knobs. Before turning on the Vibro-Deairator, always check 
that both sets of knobs are secure.

NOTE: 

Pycnometers  must  be  secured  but  NOT 

over-tightened.  Over-tightening  can  damage  the 
pycnometer or result in loss of action.

 

4.0  OPERATING INSTRUCTIONS:

NOTE: 

If  the  Vibro-Deairator  is  started  with  the 

vibration control set to LOW, the vibrator may start 
slowly, or not at all. If this occurs, rotate the knob 
counter-clockwise to a higher setting, then reduce 
the setting once the vibrator is running.

Rotate  the  vibration  control  knob  on  the  right  counter-
clockwise to its <OFF> position. Push the mode switch 
on the left to <Manual Mode> (see Figure 4). Turn the 
amplitude knob clockwise to turn on vibration.

NOTE:

 When switched to ON, the amplitude control 

goes immediately to MAXIMUM. Continue adjusting 
clockwise to minimum. 

Due  to  varying  pycnometer  and  sample  weights,  the 
operator  must  determine  the  minimum  time  and  vibra-
tion  level  necessary  to  dislodge  entrapped  air  without 
degrading the sample.

NOTE: 

Excessive vibratory action will degrade the 

sample particles by abrasion. 

Start the test by moving the mode switch to the <TIMED> 
position and pressing the <START/STOP> button on the 
timer (see below for timer set-up). The unit will run to the 

set-time and stop. The timer need not be reset unless a 
different run time is required.

This  unit  is  equipped  with  an  easy-to-operate  Gilson 
interval  count-down  timer.  The  timer  has  a  large  0.6in 
LED display and will operate in four different modes. It is 
powered by line voltage and will work on power supplies 
from 100—265 VAC, 50/60Hz, with up to 20 amps Induc-
tive or Resistive current.

NOTE:

  Operation  of  the  vibratory  motor  for  the 

SGA-5R is limited to 115V/60Hz AC. For 230V/50Hz 
operation, the TR-200 Step-Down Transformer or 
equivalent must be used.

Current timer mode is indicated by the four red LED’s on 
the timer face (see Figure 4).

A = MMSS  (99min:59sec x 1 second)
B = HHMM  (99hr:59min x 1 minute)
C = SSSS  (9999sec x 1 second)
D = MMMM  (9999min x 1 minute)

H is for hours, M for minutes, and S for seconds. To ad-
just the timer mode, press and hold both the <UP> and 
<DOWN> arrow keys at the same time until the display 
shows the mode. Once the mode letters are displayed, 
press the <UP> or <DOWN> key to change modes. Press 
the <START> key to accept new mode. 

To set the run time, press either the <UP> or <DOWN> 
arrow key. The first digit on the right hand side will flash in 
half-second intervals. Press either arrow key to adjust to 
the desired value. To enter the displayed digit and move 
to the next, press the <START/STOP> key. Once the last 
digit on the left is entered, the timer is ready to start. 

Press the <START/STOP> key to initiate the run program. 
Once running, pressing the <START/STOP> key again 
will pause the timer with the current amount of time re-
maining on screen. When allowed to time-out, the timer 
beeps and displays DONE. Press any key to continue. 
Setting  and  Mode  values  are  saved  automatically  and 
restored on power-up.

Summary of Contents for SGA-5R

Page 1: ...L Vibro Deairator SGA 5R SGA 5RT Rev 02 01 2013 PHONE 800 444 1508 P O Box 200 Lewis Center Ohio 43035 0200 FAX 800 255 5314 740 548 7298 E mail customerservice gilsonco com Website www globalgilson c...

Page 2: ...ck or damage to the unit may occur if the unit is exposed to excessive moisture 3 ALWAYS check electrical wiring for loose connections and for pinched or frayed wiring 4 ALWAYS use a properly wired th...

Page 3: ...uction 4 2 0 Unpacking Assembly 4 3 0 Set Up 4 3 1 Glass Filter Flasks 4 3 2 Aluminum Pycnometers 4 3 3 SG 15 Plastic Pycnometer 5 4 0 Operating Instructions 5 5 0 Diagrams 6 5 1 Figure 1 Glass Filter...

Page 4: ...Remove the base assembly from the box and place on a dry sturdy and level surface and connect the power cord On the rear of the unit is a power entry module with 2 amp fuses A spare fuse is located in...

Page 5: ...icles by abrasion Start the test by moving the mode switch to the TIMED position and pressing the START STOP button on the timer see below for timer set up The unit will run to the set time and stop T...

Page 6: ...e 1 Using GW 75 or GW 76 Glass Filter Flasks SGA 8 Top Adapter Top Plate Speed Knob Threaded Knob Adjustment Rod SGA 8 Rubber Base Pad Base Plate Vibration Control Timer Controller ON OFF Timed Operat...

Page 7: ...ure 2 Using SG 16 or SG 17 Aluminum Pycnometer 5 2 Figure 2 Aluminum Pycnometer SG 16 or SG 17 Aluminum Pycnometer No Adapter Required Speed Knob Top Plate Threaded Knob Adjustment Rod Base Plate Vibr...

Page 8: ...gure 3 Using SG 15 Plastic Pycnometer 5 3 Figure 3 Plastic Pycnometer SGA 7 Upper Adapter SG 15 Plastic Pycnometer SGA 7 Adapter Ring Speed Knob Top Plate Threaded Knob Adjustment Rod Base Plate Vibra...

Page 9: ...d control and a three position switch for manual or timed operation all front panel mounted for easy use Speed control allows setting of agitation to avoid stripping of asphalt The SGA 5R comes with f...

Reviews: