Gilson Company, Inc. 

Vibro-Deairator: SGA-5R & SGA-5RT

Page 4

 

1.0  INTRODUCTION:

The SGA-5R Vibro-Deairator automates sample agitation 
in specific gravity procedures such as those described 
for the AASHTO T 209 and ASTM D 2041 “Rice Test”. 
This frees laboratory personnel for other duties. The unit 
allows the use of three different types of pycnometers of 
various sizes; heavy-wall Glass Filter Flasks (Gilson 2L 
GW-75, and 4L GW-76), Aluminum Pycnometers (Gilson 
2,000g SG-16, and 1,500g SG-17), and the Gilson 6,000g 
SG-15 Plastic Pycnometer.

‡

*: DQG *: *ODVV )LOWHU )ODVNV UHTXLUH WKH

SGA-8 Adapter Set, consisting of an upper adapter 
ring and base pad.

‡

6*3ODVWLF3\FQRPHWHUUHTXLUHVWKH6*$$GDSWHU

Set, consisting of a base adapter, and upper adapter 
with foam ring.

‡

6*DQG6*$OXPLQXP3\FQRPHWHUV

DO NOT 

require adapters.

NOTE:

 ALWAYS use the correct adapter for the se-

lected pycnometer. Failure to do so can result in injury 
to the operator and damage to the pycnometer. 

NOTE:

 The SGA-5R has only been tested with the 

Gilson Pycnometers above. It CANNOT be guar-
anteed to function properly with pycnometers from 
other manufacturers.

 

2.0  UNPACKING & ASSEMBLY:

NOTE: 

Immediately upon receipt, check the shipping 

container carefully for apparent damage. If internal 
or  external  damage  is  observed  or  suspected, 
contact the freight carrier right away. Gilson can-
not be responsible for shipping damage. Save all 
shipping packaging until assured that the device is 
operating properly.

The  Vibro-Deairator  is  shipped  in  a  sturdy  cardboard 
box,  packed  in  moldable  foam.  The  top  plate,  as  well 
as two threaded height-adjustment rods with installation 
nuts, plastic and metal clamping knobs threaded on, are 
enclosed in the same box, but not attached. If the SGA-7 
or SGA-8 Adapter Sets were ordered, they will be pack-
aged separately.

Remove the base assembly from the box and place on 
a dry, sturdy and level surface and connect the power 
cord.  On  the  rear  of  the  unit  is  a  power  entry  module 
with 2 amp fuses. A spare fuse is located in the fuse tray 
should it be needed.

To install the adjustment rods, remove the nut closest to 
the end of the rod, insert the rod through the hole in the 
base plate and use two 9/16in wrenches to tighten the nuts 
and secure the rod. Repeat for the opposite side.

 

3.0  SET

-

UP:

3.1 Glass Filter-Flasks:

Place  the  rubber  base  pad  between  the  four  rubber 
bumpers on the agitator base, and then center the flask 
on the pad. Place the top adapter of the SGA-8 set over 
the neck of the flask, tilting as required to pass over the 
tubulation. Slide the top plate over the adjustment rods, 
and center the adapter between the bumpers (see Figure 
1). Position the adapter/plate assembly approximately level 
and adjust the black plastic knobs until there is a small 
gap  between  the  plate  and  the  knobs.  Drop  the  metal 
speed-knobs onto the rods and tighten so that the plate 
is pushed down slightly and contacts the bottom plastic 
knobs. Before turning on the Vibro-Deairator always check 
that both sets of knobs are secure.

NOTE: 

Pycnometers  must  be  secured  but  NOT 

over-tightened.  Over-tightening  can  damage  the 
pycnometer or result in loss of action.

3.2 Aluminum Pycnometers:

Place the pycnometer directly on the agitator base be-
tween the rubber bumpers. Slide the top plate over the 
adjustment rods, and position the rubber bumpers around 
the pycnometer (see Figure 2). Adjust the black plastic 
knobs until there is a small gap between the plate and 
the  knobs.  Drop  the  metal  speed-knobs  onto  the  rods 
and tighten so that the plate is pushed down slightly and 
contacts the bottom plastic knobs. Before turning on the 
Vibro-Deairator  always  check  that  both  sets  of  knobs 
are secure.

NOTE: 

 Pycnometers need to be secured but NOT 

over-tightened.  Over-tightening  can  damage  the 
pycnometer or result in loss of action.

Summary of Contents for SGA-5R

Page 1: ...L Vibro Deairator SGA 5R SGA 5RT Rev 02 01 2013 PHONE 800 444 1508 P O Box 200 Lewis Center Ohio 43035 0200 FAX 800 255 5314 740 548 7298 E mail customerservice gilsonco com Website www globalgilson c...

Page 2: ...ck or damage to the unit may occur if the unit is exposed to excessive moisture 3 ALWAYS check electrical wiring for loose connections and for pinched or frayed wiring 4 ALWAYS use a properly wired th...

Page 3: ...uction 4 2 0 Unpacking Assembly 4 3 0 Set Up 4 3 1 Glass Filter Flasks 4 3 2 Aluminum Pycnometers 4 3 3 SG 15 Plastic Pycnometer 5 4 0 Operating Instructions 5 5 0 Diagrams 6 5 1 Figure 1 Glass Filter...

Page 4: ...Remove the base assembly from the box and place on a dry sturdy and level surface and connect the power cord On the rear of the unit is a power entry module with 2 amp fuses A spare fuse is located in...

Page 5: ...icles by abrasion Start the test by moving the mode switch to the TIMED position and pressing the START STOP button on the timer see below for timer set up The unit will run to the set time and stop T...

Page 6: ...e 1 Using GW 75 or GW 76 Glass Filter Flasks SGA 8 Top Adapter Top Plate Speed Knob Threaded Knob Adjustment Rod SGA 8 Rubber Base Pad Base Plate Vibration Control Timer Controller ON OFF Timed Operat...

Page 7: ...ure 2 Using SG 16 or SG 17 Aluminum Pycnometer 5 2 Figure 2 Aluminum Pycnometer SG 16 or SG 17 Aluminum Pycnometer No Adapter Required Speed Knob Top Plate Threaded Knob Adjustment Rod Base Plate Vibr...

Page 8: ...gure 3 Using SG 15 Plastic Pycnometer 5 3 Figure 3 Plastic Pycnometer SGA 7 Upper Adapter SG 15 Plastic Pycnometer SGA 7 Adapter Ring Speed Knob Top Plate Threaded Knob Adjustment Rod Base Plate Vibra...

Page 9: ...d control and a three position switch for manual or timed operation all front panel mounted for easy use Speed control allows setting of agitation to avoid stripping of asphalt The SGA 5R comes with f...