QUICK GUIDE

 

SAFE ASPIRATION KIT

Cleaning and Special Care

•  The safe aspiration hand operator and accessories are fully autoclavable.
•  Clean by immersing all pieces in a solution of water and mild laboratory detergent. 

Do not use abrasives, as this may affect operation of the control valve. Rinse with 

distilled water.

•  Autoclave the unit prior to disassembly for cleaning after working with 

biohazardous materials. Refer to Image 3 for disassembly instructions.

•  Immediately flush the aspiration hand operator with ethanol after working with 

organic solvents.

•  To avoid build-up of deposits, immediately flush the aspiration hand operator with 

distilled water after working with salt-rich solutions.

•  Apply a thin coat of silicon grease as needed to the control valve. (Refer to Figure 3)

Recommended Tips

For optimum performance of the single- and 8-channel plastic tip adapters with 

ejector, we recommend using PIPETMAN® DIAMOND Tips DL10, D200, and D300.

Warranty

Gilson warrants this Safe Aspiration Kit against defects in material under normal use 

and service for 1 year from the date of purchase. This warranty is valid only if the 

components of the kit are used in the manner described in this guide and for the 

purpose for which it is designed. Gilson is not responsible for consequential damages 

resulting from the misuse or bad cleaning or decontamination of this instrument.

Forced opening of the instrument invalidates any warranty claim. Enclose with the 

returned instrument a description of the trouble that has occurred and specify which 

reagents were used. 

Please clean the instrument carefully before returning it. Never send in instruments filled 

with reagents. Returned instruments can only be checked and repaired if they have been 

carefully cleaned and decontaminated by the customer. Note that freight costs related to 

the return of the instrument will be the responsbility of the customer.

NOTICE

Installation

Insert the back of the aspiration hand operator with barbed fitting into the 

silicon tubing of the aspiration system. (Refer to Image 2)
Insert a disposable aspiration pipette, the single- or 8-channel stainless steel 

adapter directly into the port on the front of the aspiration hand operator.  

(Refer to Image 2)
For the single- or 8-channel plastic tip adapter with ejector, remove the bayonet 

cap and silicon adapter. Then screw on the plastic tip adapter with ejector.  

(Refer to Image 2)
Vacuum pressure can be controlled by depressing dark blue button on top of the 

aspiration hand operator. (Refer to Image 2)

Residual vacuum can be controlled by turning dark blue plastic screw at bottom of the 

aspiration hand operator. Turning screw one full turn clockwise shuts the valve and cuts 

off the vacuum. Turning screw counter-clockwise adjusts the constant suction level.

The safe aspiration hand operator is factory pre-set to allow a low level of constant suction.

NOTICE

NOTE

C

A

B

Summary of Contents for Safe Aspiration

Page 1: ...ip adapter with ejector Delrin acetal resin 8 channel stainless steel needle adapter Delrin acetal resin and stainless steel Four single channel stainless steel needle adapters various sizes 4 cm length 1 5 mm diameter 4 cm length 3 mm diameter 12 cm length 1 5 mm diameter 12 cm length 3 mm diameter Delrin acetal resin and stainless steel not pictured Silicon adapter for Pasteur pipette Silicon no...

Page 2: ...mination of this instrument Forced opening of the instrument invalidates any warranty claim Enclose with the returned instrument a description of the trouble that has occurred and specify which reagents were used Please clean the instrument carefully before returning it Never send in instruments filled with reagents Returned instruments can only be checked and repaired if they have been carefully ...

Page 3: ...r inoxydable Adaptateurs monocanal en acier inoxydable différentes tailles Longueur 4 cm diamètre 1 5 mm Longueur 4 cm diamètre 3 mm Longueur 12 cm diamètre 1 5 mm Longueur 12 cm diamètre 3 mm Résine d acétal Delrin et acier inoxydable non représenté Adaptateur en silicone pour pipette Pasteur Silicone non représenté Graisse silicone Silicone 3 1 2 5 6 4 Image 3 Disassembly of the hand operator Dé...

Page 4: ...raisse silicone si nécessaire sur la valve de contrôle Cf Image 3 Pointes Recommandées Pour des performances optimales de la poignée d aspiration munie de l adaptateur monocanal ou 8 canaux en plastique avec éjecteur de pointes nous recommandons l utilisation des pointes PIPETMAN DIAMOND Tips DL10 D200 et D300 disponibles dans plusieurs formats de conditionnements Garantie Gilson garantit les comp...

Reviews: