DISPENSING

Place a 96-well microplate at the front location of the left drawer

 

 

and keep it fully 

retracted in the rear position.

Move the pipetting head

 

 to the left and 

align the tips with the microplate.

• 

Pull the pipetting head slowly down 

into the wells , but do not touch the 

bottom of the wells. 

Squeeze the pipetting handle

  

slowly to 

dispense all the solution.

• 

Pull the pipetting head slowly up with 

the pipetting handle fully squeezed.

• 

Gently retract the pipetting handle.

To simplify this operation, a height adjuster

 

4

 

can be set by turning the upper screw 

so that the pipetting head may only be 
moved down to a position slightly above 
the bottom.

Pipetting 384 wells 

PLATEMASTER is able to load 384-well plates through four movements using the 
384-well plate adapter.

STEP 1:

• 

Place a 384-well microplate on the accessory and make sure that it is located at 

the first step by checking the position on the blue wheel.

• 

Aspirate the solution.

• 

Move the pipetting head over the microplate and dispense the first 96 wells.

STEP 2:

• 

Move the microplate on to the second position of the 384-well plate adapter 

using the blue wheel.

• 

Aspirate the solution. Move the pipetting head over the microplate and 

dispense the next 96 wells.

Figure 10

384-well adapter system

STEP 3:

• 

Move the microplate on to the third position of the 384-well plate adapter 

using the blue wheel.

• 

Aspirate the solution. Move the pipetting head over the microplate and 

dispense 96 wells.

STEP 4:

• 

Move the microplate on to the fourth position of the 384-well plates adapter 

using the blue wheel.

• 

Aspirate the solution. Move the pipetting head over the microplate and 

dispense the last 96 wells.

The 384-well microplate is loaded in less than one minute requiring only four liquid 
aspiration steps.

Tip Ejection

• 

Move the pipetting head over the empty rack on the right drawer.

• 

Pull the pipetting head down to put the used tips into the original rack (tip rack 

will be reused as a receptacle for tip waste).

• 

Squeeze the tip-ejector 

 at the top of the pipetting head to eject the tips into 

the rack. 

Figure 11

Tip ejection system

O

PE
R

A

TIO

N

O

PE

R

A

TIO

N

11

10

PLATEMASTER®

 

|

 

USER’S GUIDE

OPERATION

 

PLATEMASTER®

Summary of Contents for PLATEMASTER P20

Page 1: ...User s Guide EN...

Page 2: ...g master mixes for PCR qPCR Simultaneous processing of nearly all reaction steps in microplates coating starting stopping washing etc for applications such as ELISA Addition and removal of culture med...

Page 3: ...ling of the pipetting head 1 transport box and packing material Gilson original certificate of conformity PLATEMASTER P20 part number F110761 contains PLATEMASTER P20 User s guide 3 height adjusters 1...

Page 4: ...increments 2 to 220 L Volume can be adjusted in 1 L increments Temperature Range Recommended working temperature 20 C to 25 C 68 F to 77 F Dimensions 300 x 450 x 400 mm 12 x 18 x 16 Inch W x L x H Wei...

Page 5: ...a vessel containing the solution to be pipetted in the front location of the right drawer The solution should be under the pipetting head Liquids containing protein solutions and organic solvents can...

Page 6: ...icroplate and dispense the first 96 wells STEP 2 Move the microplate on to the second position of the 384 well plate adapter using the blue wheel Aspirate the solution Move the pipetting head over the...

Page 7: ...id sample or reagent Each new tip should be pre rinsed with the liquid to be pipetted Do not pipette liquids having temperatures above 70 C or below 4 C PLATEMASTER can be used between 4 C and 40 C bu...

Page 8: ...nium Use of the Lubrication Box In case of intensive use between each maintenance interval and to avoid bad tip fitting it is recommended to lightly lubricate the external O ring using the lubrication...

Page 9: ...es in Microliters Chapter 8 WARRANTY Gilson warrants this device against defects in material under normal use and service for a period of 12 months from the date of purchase This warranty shall not ap...

Page 10: ...NOTES NOTES Notes Notes 19 18 PLATEMASTER USER S GUIDE Notes PLATEMASTER...

Page 11: ...ed except as allowed under copyright laws www gilson com contactus Gilson Inc 3000 Parmenter Street PO Box 620027 Middleton WI 53562 USA Tel 608 836 1551 or 800 445 7661 F 608 831 4451 Gilson S A S 19...

Reviews: