Chapter 3

OPERATION

Description

The pipetting head moves both vertically and horizontally. Vertical movements 

are made to fit tips and pipette liquid. Horizontal movements are made to 

transfer liquids. This movement is possible only when the head is fully raised.
Two drawers can hold up to three different microplates and can be used at the 

same time.

General View

The figure below shows a general view of a PLATEMASTER.

   

Pipetting head

   

Volume adjustment wheel

   

Volume display 

   

Tip fitting handle

3

2

1

6

5

4

8

7

Figure 7

PLATEMASTER

® 

parts

Height Adjuster

The descent of the pipetting head and the 

position of the tips can be adjusted by means 
of three height adjusters 

.

This will prevent the tips from touching 

the bottom of the microplate and can be 

adjusted for use with shallow- or deep-well 

microplates.

Figure 8

PLATEMASTER

®

 lock

Figure 9

Three height adjusters positions

Technical Data

MODEL

PLATEMASTER P20

PLATEMASTER P220

Part Number

F110761

F110762

Volume Range

0.5 to 20 µL

Volume can be adjusted 

in 0.1 µL increments

2 to 220 µL

Volume can be adjusted 

in 1 µL increments

Temperature 

Range

Recommended  working  temperature:                     

20°C to 25°C / 68°F to 77°F 

Dimensions

300 x 450 x 400 mm / 12 x 18 x 16 Inch (W x L x H)

Weight

approx. 16 kg / approx. 35 lbs.

   

Pipetting handle

  

 Tip ejector

   

Movable drawers

   

Pipetting head height adjuster

9

   

Pin-plate

Moving PLATEMASTER

Whether placed on the lab bench, under a fume hood, 

in a cold room, or used in the field, the PLATEMASTER 
is completely portable. The PLATEMASTER includes a 
locking device to lock the head when the instrument is 

being moved.
After each use of the PLATEMASTER, remember to lock 

the pipetting head to prepare for transport.
To lock the pipetting head, pull the black part on the 

left side of the PLATEMASTER, lower the head, and turn 

the black part until the notch is in the slot. Let the head 

slowly ride up and be locked.
To unlock the pipetting head, pull the black part and turn it. Once the pipetting head 

is released, check proper movement of the device.

9

Never move the device without locking the pipetting head. 

Damages caused by movement without locking the device will not 

be covered by the warranty.

O

PE
R

A

TIO

N

IN

ST

A

LL

A

TIO

N

7

6

PLATEMASTER®

 

|

 

USER’S GUIDE

INSTALLATION

 

PLATEMASTER®

Summary of Contents for PLATEMASTER P20

Page 1: ...User s Guide EN...

Page 2: ...g master mixes for PCR qPCR Simultaneous processing of nearly all reaction steps in microplates coating starting stopping washing etc for applications such as ELISA Addition and removal of culture med...

Page 3: ...ling of the pipetting head 1 transport box and packing material Gilson original certificate of conformity PLATEMASTER P20 part number F110761 contains PLATEMASTER P20 User s guide 3 height adjusters 1...

Page 4: ...increments 2 to 220 L Volume can be adjusted in 1 L increments Temperature Range Recommended working temperature 20 C to 25 C 68 F to 77 F Dimensions 300 x 450 x 400 mm 12 x 18 x 16 Inch W x L x H Wei...

Page 5: ...a vessel containing the solution to be pipetted in the front location of the right drawer The solution should be under the pipetting head Liquids containing protein solutions and organic solvents can...

Page 6: ...icroplate and dispense the first 96 wells STEP 2 Move the microplate on to the second position of the 384 well plate adapter using the blue wheel Aspirate the solution Move the pipetting head over the...

Page 7: ...id sample or reagent Each new tip should be pre rinsed with the liquid to be pipetted Do not pipette liquids having temperatures above 70 C or below 4 C PLATEMASTER can be used between 4 C and 40 C bu...

Page 8: ...nium Use of the Lubrication Box In case of intensive use between each maintenance interval and to avoid bad tip fitting it is recommended to lightly lubricate the external O ring using the lubrication...

Page 9: ...es in Microliters Chapter 8 WARRANTY Gilson warrants this device against defects in material under normal use and service for a period of 12 months from the date of purchase This warranty shall not ap...

Page 10: ...NOTES NOTES Notes Notes 19 18 PLATEMASTER USER S GUIDE Notes PLATEMASTER...

Page 11: ...ed except as allowed under copyright laws www gilson com contactus Gilson Inc 3000 Parmenter Street PO Box 620027 Middleton WI 53562 USA Tel 608 836 1551 or 800 445 7661 F 608 831 4451 Gilson S A S 19...

Reviews: