INSTALLATION QUALIFICATION/OPERATIONAL QUALIFICATION PROCEDURES
PIPETMAX® 268
LT255520-05 | Page 1 of 7
These installation qualification and operational qualification (IQ/OQ) procedures are intended for
end users who are implementing good laboratory practice (GLP) requirements and can be readily
incorporated into proprietary standard operating procedures (SOPs).
1. Instrument Identification
Manufacturer
Gilson, Inc.
3000 Parmenter Street
PO Box 620027
Middleton, WI 53562-0027 USA
Instrument Name
Record Serial Number
Serial Number Locations
(Compare to ensure they match)
PIPETMAX®
●
Shipping label on the shipping
container
●
Inside panel of the right
support on the instrument
Record Serial Number(s)
Pipette heads are required, but ordered
separately. Record the serial number(s)
below, if applicable.
Serial Number Locations
(Compare to ensure they match)
MAX8x20 Pipette Head(s)
●
Shipping envelope
●
Document supplied with the
pipette head
●
Pipette head
MAX8x200 Pipette Head(s)
●
Shipping envelope
●
Document supplied with the
pipette head
●
Pipette head
MAX1000 Pipette Head(s)
●
Shipping envelope
●
Document supplied with the
pipette head
●
Pipette head
MAX4x1200 Pipette Head(s)
●
Shipping envelope
●
Document supplied with the
pipette head
●
Pipette head