background image

6

QUICK GUIDE 

|

 

PIPETMAN® L

Setting the Volume

The volume of liquid to be aspirated is set using the volume display. 

The dial colors are either black or red to indicate the position of the 

decimal point, depending on the model (refer to Figure 3).

Figure 3

Dial colors by model

Model

Color of Volumeter Numbers

BLACK

RED

INCREMENT

P2L

µL

0.01 µL

0.002 µL

P10L to P20L – 

P8X10L - P12X10L

µL

0.1 µL

0.02 µL

P100L – P200L 

P8X200L - P12X200L 

P8X300L - P12X300L

µL

-

0.2 µL

P8X200LVR - 

P12X200LVR 

P8X300LVR - 

P12X300LVR

 

µL

-

0.2 µL

P1000L

0.01 mL mL

0.002 mL

P8x1200L - P12x1200L 0.01 mL mL

0.002 mL

P5000L

0.01 mL mL

0.002 mL

P10mLL

mL

0.1 mL

0.02 mL

PIPETMAN® L MULTICHANNEL MODELS

Vol.

(µL)

Maximum Permissible Errors

Gilson

ISO 8655

Systematic 

Error 

(µL) 

Random 

Error 

(µL)

Systematic 

Error 

(%) 

Random 

Error 

(%CV*)

Systematic 

Error 

(µL) 

Random 

Error 

(µL)

Systematic 

Error 

(%) 

Random 

Error 

(%CV*)

P8X10L (P/N FA10013) AND P12X10L (P/N FA10014) DL10 TIPS

0.5 

10 

± 0.08 

± 0.08 

± 0.20 

± 0.20

 0.04 

 0.05 

 0.10 

 0.10

± 16.0

± 8.0

± 4.0

± 2.0

 

8.0

 

5.0

 

2.0

 

1.0

± 0.24 

± 0.24 

± 0.24 

± 0.24

 0.16 

 0.16 

 0.16 

 0.16 

± 48.0

± 24.0

± 4.8

± 2.4

 

32.0

 

16.0

 

3.2

 

1.6

P8X20L (P/N FA10009) AND P12X20L (P/N FA10010) WITH DL10 TIPS

10 

20

± 0.10 

± 0.20 

± 0.40

 0.08 

 0.10 

 0.20

± 5.0

± 2.0

± 2.0

 

4.0

 

1.0

 

1.0

± 0.40 

± 0.40 

± 0.40

 0.20 

 0.20 

 0.20

± 20.0

± 4.0

± 2.0

 

10.0

 

2.0

 

1.0

P8X200L (P/N FA10011) AND P12X200L (P/N FA10012) WITH D200 TIPS

20 

100 

200

± 0.50 

± 1.00 

± 2.00

 0.25 

 0.40 

 0.50

± 2.5

± 1.0

± 1.0

 

1.25

 

0.4

 

0.25

± 3.2 

± 3.2 

± 3.2

 1.2 

 1.2 

 1.2 

± 16.0

± 3.2

± 1.6

 

6.0

 

1.2

 

0.6

P8X200LVR (P/N FA10035) AND P12X200LVR (P/N FA10036) WITH D200 TIPS

20 

100 

200

± 0.50 

± 1.00 

± 2.00

 0.25 

 0.40 

 0.50

± 2.5

± 1.0

± 1.0

 

1.25

 

0.4

 

0.25

± 3.2 

± 3.2 

± 3.2

 1.2 

 1.2 

 1.2

± 16.0

± 3.2

± 1.6

 

6.0

 

1.2

 

0.6

P8X300L (P/N FA10015) AND P12X300L (P/N FA10016) WITH D300 TIPS

20 

30 

150 

300

± 1.00 

± 1.00 

± 1.50 

± 3.00

 0.35 

 0.35 

 0.60 

 1.00

± 5.0

± 3.33

± 1.0

± 1.0

 

1.75

 

1.17

 

0.4

 

0.33

± 8.0 

± 8.0 

± 8.0 

± 8.0

 3.0 

 3.0 

 3.0 

 3.0 

± 40.0

± 26.67

± 5.33

± 2.67

 

15.0

 

10.0

 

2.0

 

1.0

P8X300LVR (P/N FA10037) AND P12X300LVR (P/N FA10038) WITH D300 TIPS

20 

30 

150 

300

± 1.00 

± 1.00 

± 1.50 

± 3.00

 0.35 

 0.35 

 0.60 

 1.00

± 5.0

± 3.33

± 1.0

± 1.0

 

1.75

 

1.17

 

0.4

 

0.33

± 4.8

± 4.8

± 4.8

± 4.8

 1.8

 1.8

 1.8

 1.8

± 24.0

± 16.0

± 3.2

± 1.6

 

9.0

 

6.0

 

1.2

 

0.6

P8X1200L (P/N FA10039) AND P12X1200L (P/N FA10040) WITH D1200 TIPS

100

120

600

1200

± 6.0

± 6.0

± 8.0

± 10.0

 0.90

 0.90

 1.20

 1.50

± 6.0

± 5.0

± 1.3

± 0.8

 

0.9

 

0.8

 

0.2

 

0.1

± 32

± 32

± 32

± 32

 

12

 

12

 

12

 

12

± 32.0

± 26.7

± 5.3

± 2.7

 

12.0

 

10.0

 

2.0

 

1.0

*CV means Coefficient of Variation

1

5

 

P2L

1.25 µL

2

0

7

5

 

P10L

7.5 µL

1

2

5

 

P20L

12.5 µL

 

P100L

0

7

5

75 µL

 

P200L

1

2

5

125 µL

 

P1000L

0

7

5

0.75 mL

 

P5000L

1

2

5

1.25 mL

 

P10mLL

0

7

5

7.5 mL

0

7

5

 

Px10L

7.5 µL

1

2

5

 

Px20L

12.5 µL

 

Px200L

Px200LVR

1

2

5

125 µL

 

Px300L

Px300LVR

1

2

5

125 µL

 

Px1200L

0

7

5

0.75 µL

Multichannel models

Single channel models

Summary of Contents for PIPETMAN L

Page 1: ...Quick Guide EN A U TOCLAVABL E FULLY VARIABLE VOLUME MODELS...

Page 2: ...Certificate of Conformity including barcode sticker Tip ejector adapter equipped on P2L P10L Bag of ten filters only for P5000L and P10mLL Multichannel Models PIPETMAN L Quick Guide Safety bag Certifi...

Page 3: ...ngle Channel Models 2 5 1 2 3 4 1 3 3 2 1 4 4 5 7 8 8 11 11 11 7 8 10 9 6 5 7 1 Color coded push button 2 Thumbwheel 3 Tip ejector button 4 Handle 5 Connecting nut 6 Ejector clip 7 Tip holder s 8 Tip...

Page 4: ...0 TIPS 5 0 050 0 025 1 0 0 5 0 125 0 075 2 5 1 5 F10L P N FA10020 WITH D10 AND DL10 TIPS 10 0 06 0 03 0 6 0 3 0 12 0 08 1 2 0 8 F20L P N FA10021 WITH D200 TIPS 20 0 10 0 05 0 5 0 25 0 20 0 10 1 0 0 5...

Page 5: ...0 3 0 2 0 2 0 2 0 1 0 1 0 1 10 0 2 0 1 0 5 0 1 0 0 5 P100L P N FA10004M AND FA10004P WITH D200 TIPS 10 50 100 0 35 0 40 0 80 0 10 0 12 0 15 3 5 0 8 0 8 1 0 0 24 0 15 0 8 0 8 0 8 0 3 0 3 0 3 8 0 1 6 0...

Page 6: ...20 0 20 20 0 4 0 2 0 10 0 2 0 1 0 P8X200L P N FA10011 AND P12X200L P N FA10012 WITH D200 TIPS 20 100 200 0 50 1 00 2 00 0 25 0 40 0 50 2 5 1 0 1 0 1 25 0 4 0 25 3 2 3 2 3 2 1 2 1 2 1 2 16 0 3 2 1 6 6...

Page 7: ...key in either direction equals 0 05 L F1L to F2L 0 2 L F5L to F10L 0 5 L F20L to F25L 2 0 L F50L to F100L 5 1 L F200L to F250L 20 L F300L to F1000L 102 L F5000L The cover is graduated in tenths of th...

Page 8: ...MAN DIAMOND Tip Compatibility for multichannel models P8x10L P12x10L D10 DL10 DF10ST DFL10ST P8x20L P12x20L DL10 D200 DFL10ST DF30ST P8x200L P12x200L D200 D300 DF200ST DF300ST P8x200LVR P12x200LVR D20...

Page 9: ...11 To fit a tip ejector extension 1 Slide the extension over the tip holder 2 Push the extension firmly onto the end of the tip ejector until it clicks into place To remove a tip ejector extension 1...

Page 10: ...of your thumb for more comfort Pre Wet the Tips Pre wetting the tips before pipetting helps prepare the tips for the best pipetting performance Ideally the pre wet includes both immersing the tip in...

Page 11: ...ecommendations given in the user s guide including but not limited to pipettes which have been subjected to physical damage improper handling or spillage or exposure to any corrosive environment This...

Page 12: ...gilson com contactus LT801634 B 2021 Gilson Inc All rights reserved...