background image

PIPETMAN

®

 F

ENGLISH

Other languages available on 
our website: www.gilson.com.

Summary of Contents for PIPETMAN F Series

Page 1: ...PIPETMAN F ENGLISH Other languages available on our website www gilson com...

Page 2: ...nd disposable tipsmeannoneedtohandle tips protectingtheuserfrom contamination The tip holder the connect ing nut and the tip ejector are autoclavable for 20 minutes at 121 C and 0 1 MPa pressure CONTE...

Page 3: ...ing genuine Gilson Diamond Tips The data given in the table conform to the ISO 8655 2 Standard Maximum Permissible Errors Model Volume Gilson ISO 8655 Reference L Systematic Random Systematic Random e...

Page 4: ...rements for each test volume 3 OPERATION Place a tip on the tip holder Press thetiponfirmlyusingaslighttwist ing motion to ensure a positive airtight seal See Chapter 6 for the appropriate tip Always...

Page 5: ...balances The nominal value of the Pipetman F may be slightly adjusted to compensate for liquids of different density or viscosity Adjustment of the factory calibration must only be performed using a s...

Page 6: ...ion tool must be turned 3 units until the display reads B 0 2 It is possible to alter the volume by more than one full turn in the counterclockwise direction In this case remember to readjust the same...

Page 7: ...ty control is maintained throughout the manufacturing process Diamond Tips are used to calibrate Pipetman F therefore for optimum performance you are stronglyadvisedtouseGilson sDiamondTipswith your P...

Page 8: ...k of precision Theymaybeautoclavedat121 Cfor20minutes at 0 1 MPa Diamond Filter Tips Filter tips are used when sample to sample sample to pipette or sample to operator con tamination must be avoided G...

Page 9: ...ent this Press and release the push button slowly and smoothly Never turn the pipette upside down Never lay the pipette on its side when there is liquid in the tip Use a dedicated stand or hanger Gils...

Page 10: ...holder using distilled water Improper reassembly Disassemble the pipette and reassemble the pipette correctly If there is liquid in the tip holder clean the pipette as follows Removethetipejector Unsc...

Page 11: ...require re calibration of the Pipetman F We recommend thatthisworkiscarriedoutby your Gilson representative Reference Number Item F2 to F20 F25 to F100 F200 F250 to F1000 F5000 Piston assembly F12384...

Page 12: ...12 PIPETMAN F...

Page 13: ...irements of the following European Directives Villiers le Bel September 14th 2005 A El Sayed H Ledorze General Manager Quality Manager GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 Villiers le...

Page 14: ...14 PIPETMAN F NOTES...

Page 15: ...15 ENGLISH...

Page 16: ...3562 0027 USA Telephone 1 800 445 7661 or 1 608 836 1551 Fax 1 608 831 4451 Gilson SAS 19 avenue des Entrepreneurs B P 145 95400 Villiers le Bel France Telephone 33 1 34 29 50 00 Fax 33 1 34 29 50 20...

Reviews: