background image

6

PIPETMAN

®

 F

the  push-button  until  notches  fit  into 
the  grooves  in  the  calibration  sleeve 
(Figure ), then turn the key: 

•  clockwise to decrease the volume;

•  counterclockwise to increase the 

volume;  go  above  the  desired 
value, then decrease the volume 
to reach the required setting.

Example: When using a Pipetman F100 to aspirate a 
particularly viscous solution, you may determine gravi-
metrically that the volume delivered is 98.8 µL, and the 
display reads ‘E 0.2’. Therefore, you wish to increase the 
volume dispensed by 1.2 µL. As the interval between 
each letter on the calibration display represents 0.4 µL 
for the F100 (1/10 of 4 µL), the calibration tool must be 
turned 3 units until the display reads ‘B 0.2’.

It is possible to alter the volume by more than 
one  full  turn  in  the  counterclockwise  direction. 
In  this  case,  remember  to  readjust  the  same 
number of turns when it is necessary to return to 
the original setting.

Check the new volume gravimetrically. If the vol-
umes delivered are still not sufficiently close to 
the desired value, make another slight adjustment 
in the relevant direction. 

Be sure to change tips 

between volume setting adjustments and to 
pre-rinse each new tip.

When the Pipetman F is readjusted 
to compensate for a particular solu-
tion, the original setting (ex. 0.) can 
be noted on the self-adhesive labels 
supplied (Figure 4). 

Liquids of high density, vapor pressure, or viscosity 
are best pipetted by the Gilson Microman

®

 positive 

displacement pipette.

Figure 3

Figure 4

Summary of Contents for PIPETMAN F Series

Page 1: ...PIPETMAN F ENGLISH Other languages available on our website www gilson com...

Page 2: ...nd disposable tipsmeannoneedtohandle tips protectingtheuserfrom contamination The tip holder the connect ing nut and the tip ejector are autoclavable for 20 minutes at 121 C and 0 1 MPa pressure CONTE...

Page 3: ...ing genuine Gilson Diamond Tips The data given in the table conform to the ISO 8655 2 Standard Maximum Permissible Errors Model Volume Gilson ISO 8655 Reference L Systematic Random Systematic Random e...

Page 4: ...rements for each test volume 3 OPERATION Place a tip on the tip holder Press thetiponfirmlyusingaslighttwist ing motion to ensure a positive airtight seal See Chapter 6 for the appropriate tip Always...

Page 5: ...balances The nominal value of the Pipetman F may be slightly adjusted to compensate for liquids of different density or viscosity Adjustment of the factory calibration must only be performed using a s...

Page 6: ...ion tool must be turned 3 units until the display reads B 0 2 It is possible to alter the volume by more than one full turn in the counterclockwise direction In this case remember to readjust the same...

Page 7: ...ty control is maintained throughout the manufacturing process Diamond Tips are used to calibrate Pipetman F therefore for optimum performance you are stronglyadvisedtouseGilson sDiamondTipswith your P...

Page 8: ...k of precision Theymaybeautoclavedat121 Cfor20minutes at 0 1 MPa Diamond Filter Tips Filter tips are used when sample to sample sample to pipette or sample to operator con tamination must be avoided G...

Page 9: ...ent this Press and release the push button slowly and smoothly Never turn the pipette upside down Never lay the pipette on its side when there is liquid in the tip Use a dedicated stand or hanger Gils...

Page 10: ...holder using distilled water Improper reassembly Disassemble the pipette and reassemble the pipette correctly If there is liquid in the tip holder clean the pipette as follows Removethetipejector Unsc...

Page 11: ...require re calibration of the Pipetman F We recommend thatthisworkiscarriedoutby your Gilson representative Reference Number Item F2 to F20 F25 to F100 F200 F250 to F1000 F5000 Piston assembly F12384...

Page 12: ...12 PIPETMAN F...

Page 13: ...irements of the following European Directives Villiers le Bel September 14th 2005 A El Sayed H Ledorze General Manager Quality Manager GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 Villiers le...

Page 14: ...14 PIPETMAN F NOTES...

Page 15: ...15 ENGLISH...

Page 16: ...3562 0027 USA Telephone 1 800 445 7661 or 1 608 836 1551 Fax 1 608 831 4451 Gilson SAS 19 avenue des Entrepreneurs B P 145 95400 Villiers le Bel France Telephone 33 1 34 29 50 00 Fax 33 1 34 29 50 20...

Reviews: