MACROMAN

TM

 

|

 

USER’S GUIDE

OPERATION

 

MACROMAN

TM

Charging MACROMAN with the Desk Stand

To charge the MACROMAN, plug the desk stand into 

an electrical outlet and then place MACROMAN into 

the stand.  

Observe the following:

 

When MACROMAN is properly seated, the 

battery icon will indicate the charging status.

 

The battery display shows between 1 to 6 bars 

in sequence from empty to full. Allow the battery 

to charge fully during 6 hours before using your 

MACROMAN for the first time.

 

When MACROMAN is completely charged, the 

battery icon shows as full (solid bars).  

(Figure 5)

Using MACROMAN 

Inserting the Pipette

MACROMAN accommodates pipette capacities 

from 1 mL to 100 mL.

Pipettes > 2 mL

MACROMAN comes assembled 

with the adaptor set to accept larger pipettes.

Pipettes 

 2 mL

:

1.  Loosen then remove the nosepiece.
2.  Remove the pipette adaptor and then flip it so 

the small hole faces down towards the pipette. 

 

 

 

 

3.   Reattach the nosepiece.
4.  Once the adaptor is set for your desired pipette size, carefully insert 

the pipette into the nosepiece, providing a secure fit.

Setting the Flow Rate 

 

Adjust the maximum speed by 

turning the thumbwheel until 

desired speed on display is 

reached.

 

When pressing either trigger, 

the speedometer reflects the 

speed up to the set thumbwheel 

position.

O

PE
R

A

TIO

N

When you flip the pipette adaptor, it is 
normal for it to be loose. Airtightness is 
maintained.

Figure 8

Rotate the pipette adaptor 

to accommodate different 

types of pipettes

Left position 

Low

Right position 

High

MASTER

Figure 9

Adjust speed with the thumbwheel to set the 

flow rate

Figure 7

MACROMAN

TM

 on its desk 

stand

Figure 6

Bottom view of the desk 

stand

O

PE

R

A

TIO

N

For Drip Mode

Set thumbwheel all the way to the left: speedometer 

disappears and drip icon flashes three times.
The pressure sensitive dispense trigger allows you  

to control the flow speed.

Filling and Dispensing

Aspirating liquid

1.  Immerse the pipette into the liquid.
2.  Press the Aspirate (upper) trigger until the pipette is filled just above the 

desired mark. 

3.  Press the Dispense (low) trigger slowly until the meniscus is exactly level 

with the desired mark.

Dispensing liquid

During blow-out, liquid is dispensed with the aid of the pump.

1.  Place the pipette against the wall of the vessel.
2.  Press the Dispense (low) trigger completely.

For drip mode 

 

Aspirate is on low motor speed. 

 

Press the Dispense (low) trigger to release the liquid drop by drop. 

Liquid dispensing takes place 

only using atmospheric pressure.

Using the Side Stand 

 

Use the side stand to safely 

place the MACROMAN and its 

pipette on a lab bench. 

 

Turn the stand to the left or to 

the right to avoid the pipette 

touching the lab bench.

 

Sleep Mode

MACROMAN goes into sleep mode after five minutes.  

Press any button to reactivate the device.

Figure 10

Flow rate for drip mode

Figure 11

MACROMAN

TM

 laying on its side stand

Do not use the maximum speed with low volume 
pipettes (

 2 mL). At high speed, the liquid will 

damage your pipette controller.

NOTICE

Summary of Contents for macroman

Page 1: ...User s Guide EN...

Page 2: ...mation 6 OPERATION 7 Operating Conditions 7 Charging MACROMAN 7 Using MACROMAN 8 MAINTENANCE CLEANING 10 Safety and User Precautions 10 Replacing the Battery 10 Replacing the Filter 11 Cleaning 11 TRO...

Page 3: ...ease make sure to read the safety and operating instructions before use Also when performing the pipetting function observe all laboratory safety guidelines including the use of protective clothing sa...

Page 4: ...e you can drop off your waste equipment for recycling please contact your local dealer from whom you originally purchased the product or your local council By doing so you will help conserve natural r...

Page 5: ...Rotate the pipette adaptor to accommodate different types of pipettes Left position Low Right position High MASTER Figure 9 Adjust speed with the thumbwheel to set the flow rate Figure 7 MACROMANTM o...

Page 6: ...e 12 MACROMANTM battery A C B Do not force the connector It only fits one way There should be a slight click when the cable connection is inserted properly A C B It is imperative to dispose of a used...

Page 7: ...ng Table SYMPTOM POSSIBLE CAUSE ACTION The serological pipette falls out the holding force of the pipette is too small or tilts to the side too much The pipette adaptor is dirty or wet Take out the pi...

Page 8: ...or replacement will be provided by your local Gilson service center Freight shipping the unit to Gilson will be the responsibility of the purchaser There are no other warranties or guarantees implied...